High-Frequency Ultrasound for Assessment of Peri-Implant Bone Thickness.

J Clin Med

Department of Prosthodontics and Biomaterials, Centre for Implantology, Medical School of the RWTH Aachen University, Pauwelsstrasse 30, 52074 Aachen, Germany.

Published: September 2019

Purpose: The aim of this study was to evaluate the accuracy of high-frequency ultrasound (HFUS) for measurement of bone thickness surrounding dental implants.

Methods: Eight porcine bone samples containing dental implants were scanned by a HFUS scanner and compared using cone-beam computed tomography (CBCT) and an optical scanner. Bone thickness was measured in the buccolingual region of dental implants in 10 points distributed between the platform and apical portion of the implant.

Results: The mean measurement error for the ultrasound method was 0.11 mm, whereas CBCT showed a measurement error of 0.20 mm. For both devices, the maximal measurement error was 0.28 mm.

Conclusion: Within the simulated limited conditions of this study, high-frequency ultrasound, with optical scanning used as a reference, presented higher accuracy in comparison to CBCT, and seems to be a promising tool for measuring peri-implant bone.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6832403PMC
http://dx.doi.org/10.3390/jcm8101539DOI Listing

Publication Analysis

Top Keywords

high-frequency ultrasound
12
bone thickness
12
measurement error
12
peri-implant bone
8
dental implants
8
bone
5
ultrasound assessment
4
assessment peri-implant
4
thickness purpose
4
purpose aim
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!