Purpose: TRASE uses phase gradients in the RF transmit field to encode MRI data. A highly efficient twisted solenoid coil was proposed recently for TRASE imaging for transverse geometries. This novel coil can be rotated to generate a phase gradient in any transverse direction, therefore, combining two such coils would double k-space coverage for single-axis encoding, resulting in higher spatial resolution. However, the strong inductive coupling between a pair of coaxial twisted solenoids must be overcome.

Methods: Here, we demonstrate that two concentric twisted solenoids, designed using previously described Biot-Savart calculations, can be geometrically decoupled by attaching to each a regular solenoid in series. The regular solenoid geometry resulting in minimization of mutual inductance was determined from simulations using the FastHenry2 tool. The effects on TRASE encoding performance due to the regular solenoids were assessed from simulations and experiments.

Results: The maximum resulting magnitude and phase distortions were 3.7% and , while a good isolation dB between the coil pair was obtained. TRASE experiments confirmed the double k-space coverage, and achieved a rapid spin echo train with 128 k-space points collected within 80 ms, allowing short samples to be accurately imaged.

Conclusions: This study demonstrates that a pair of twisted solenoid phase gradient RF coils can be geometrically decoupled. Advantages over active PIN diode decoupling include faster switching, lower hardware complexity, and scalability.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.28003DOI Listing

Publication Analysis

Top Keywords

geometrically decoupled
12
twisted solenoid
12
phase gradient
8
double k-space
8
k-space coverage
8
twisted solenoids
8
regular solenoid
8
twisted
5
solenoid
5
trase
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!