This article presents a new configurable wireless sensor system. The system is used to perform amperometric measurements and send the measurement data to a handheld reader using a wireless transponder interface. The two-chip sensor system was implemented in a 0.35 μm CMOS technology. The system consists of an integrated nano-potentiostat that performs the actual measurements and an ISO 18000-3 compliant frontend that enables wireless telemetric data transmission and powering of the entire sensor system. The system was manufactured in combination with a chronoamperometric glucose sensor which allows the measurement of the glucose content in tear fluid and thus a non-invasive determination of the blood sugar level. For a range of sensor currents from 0.1 μA to 10 μA, the potentiostat achieved an accuracy of better than 5 % with a total power dissipation of less than 600 μW. With the realized antenna geometry a wireless communication distance of more than 7 cm has been achieved.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806625PMC
http://dx.doi.org/10.3390/s19194110DOI Listing

Publication Analysis

Top Keywords

sensor system
16
configurable wireless
8
wireless sensor
8
iso 18000-3
8
system system
8
system
7
sensor
6
system biomedical
4
biomedical applications
4
applications iso
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!