Background: Most manufacturer manuals do not verify the use of dual energy X-ray absorptiometry for body composition analysis in subjects with a metal implant. This study aimed to quantify the effects of a spinal implant on body composition, and to determine whether unadjusted lean mass estimates are valid for patients with a spinal implant.

Methods: A total of 30 healthy subjects were recruited. Three consecutive scans were performed for each participant, one with and two without extraneous spinal implant, without repositioning between scans. Lean, fat and bone estimates in the total body, trunk and limb were measured.

Results: Precision errors for all total and regional body compositions were within the recommended ranges. Bone masses in the trunk and total body were significantly increased with spinal implant, and the increases exceeded the least significant change. For total and regional lean and fat estimates, the measurements between subjects with and without metal implants were in substantial to almost perfect agreement and the differences were not significant and did not exceed the least significant change.

Conclusions: Spinal metal artifacts significantly increased the total body and trunk bone mass but the differences in lean- and fat-related estimates at total and regional body levels and all estimates in the extremity remained within the clinical acceptable range. Thus, a spinal implant may not compromise screening of patients for fat and lean masses using dual energy X-ray absorptiometry. Application of image reconstruction or a filtering algorithm may help reduce the effect of metallic artifacts and further study is needed.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6752773PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0222758PLOS

Publication Analysis

Top Keywords

spinal implant
20
dual energy
12
energy x-ray
12
x-ray absorptiometry
12
body composition
12
total body
12
total regional
12
body
8
absorptiometry body
8
subjects metal
8

Similar Publications

Author Correction: Discovering CRISPR-Cas system with self-processing pre-crRNA capability by foundation models.

Nat Commun

January 2025

State Key Laboratory of Cardiology and Medical Innovation Center, Shanghai East Hospital, Frontier Science Center for Stem Cell Research, Bioinformatics Department, School of Life Sciences and Technology, Tongji University, Shanghai, China.

View Article and Find Full Text PDF

Review of surgical methods for stabilizing the lumbosacral spine in dogs with chronic L7-S1 instability.

Vet Res Commun

January 2025

Department of Biomaterials and Medical Devices Engineering, Faculty of Biomedical Engineering, Silesian University of Technology, Roosevelta 40, Zabrze, 41-800, Poland.

Chronic instability at the lumbosacral junction, particularly between the L7 vertebra and the sacral bone, presents significant challenges in veterinary orthopedics, especially for large breed dogs. This condition frequently results in severe pain, neurological deficits, and mobility impairments, prompting the development of various surgical techniques aimed at effectively stabilizing the affected area. A critical evaluation of the literature on surgical stabilization of the lumbosacral spine in dogs reveals the clinical applications, outcomes, and future directions in veterinary spinal surgery.

View Article and Find Full Text PDF

Clinical applications of 3D printing in spine surgery: a systematic review.

Eur Spine J

January 2025

In Silico Biomechanics Laboratory, National Center for Spinal Disorders, Buda Health Center, Budapest, Hungary.

Purpose: The objective of this systematic review is to present a comprehensive summary of existing research on the use of 3D printing in spinal surgery.

Methods: The researchers conducted a thorough search of four digital databases (PubMed, Web of Science, Scopus, and Embase) to identify relevant studies published between January 1999 and December 2022. The review focused on various aspects, including the types of objects printed, clinical applications, clinical outcomes, time and cost considerations, 3D printing materials, location of 3D printing, and technologies utilized.

View Article and Find Full Text PDF

Thoracic vertebral body erosion due to a perianeurysmal outpouching lesion after thoracic endovascular aortic repair: a case report and literature review.

BMC Musculoskelet Disord

January 2025

Spine Center and Department of Orthopedic Surgery, Inje University Sanggye Paik Hospital, College of Medicine, Inje University, 1342, Dongil-Ro, Seoul, Nowon-Gu, 01757, Republic of Korea.

Background: The safety of endovascular treatment, such as thoracic endovascular aortic repair (TEVAR), for a descending thoracic aortic aneurysm has been well-established, with a reported low postoperative mortality rate but higher incidences of long-term complications such as endo-leakage, device failure, and aneurysm-related death. Based on this, we report the first case of massive thoracic vertebral body erosion due to a perianeurysmal outpouching lesion after TEVAR.

Case Presentation: A 77-year-old female with a history of TEVAR due to descending thoracic aortic arch aneurysm 4 years ago was referred from the cardiovascular clinic to the spine center.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!