Purpose: To investigate intraocular pressure (IOP) readings by non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT) for patients with different degrees of bilateral tearing.
Methods: In this study, we reviewed the medical charts of patients complaining of different degrees of bilateral tearing. The tear meniscus height (TMH) and IOP with NCT and GAT were measured. In each patient, a comparison of IOP readings between the eye with lower TMH and the contralateral eye with higher TMH was evaluated. The TMH was graded as follows: grade 1 (low): TMH < 0.2 mm; grade 2 (moderate): 0.2 mm ≤ TMH < 0.6 mm; grade 3 (high): TMH ≥ 0.6 mm. Subsequently, a comparison of IOP readings among eyes with low, moderate, and high TMH was also performed.
Results: A total of 120 eyes of 60 patients were enrolled. When comparing the two eyes of a patient, the eye with higher TMH showed higher NCT readings and larger difference in IOP readings between the two tonometries than the eye with lower TMH (P < 0.001 and P < 0.001, respectively). When TMH was classified into grades according to the degree, the high TMH eyes showed higher NCT readings than did the low and moderate TMH eyes (P < 0.001 and P = 0.001, respectively). In addition, the high TMH eyes showed a larger difference in IOP readings between the two tonometries than did the low and moderate TMH eyes (P < 0.001 and P < 0.001, respectively).
Conclusion: Eyes with higher TMH showed higher NCT readings and a larger difference in IOP between the two tonometries (NCT and GAT) than those with lower TMH. In patients with tearing, the NCT value may be inaccurate, so it is necessary to measure the GAT.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6746371 | PMC |
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0222652 | PLOS |
J Clin Med
November 2024
2nd Department of Ophthalmology, School of Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki, 54124 Thessaloniki, Greece.
Intraocular pressure (IOP) readings using three different methods (Goldmann applanation tonometry (GAT), Corvis ST, and iCare) were compared in patients who underwent penetrating keratoplasty (PK). : An observational cross-sectional study with prospective recruitment of patients was conducted. IOP measurements were acquired using GAT, iCare, and Corvis (including both uncorrected IOP (CVS-IOP) and biomechanical IOP (bIOP)), and the agreement among methods was analyzed using Bland-Altman plots.
View Article and Find Full Text PDFOphthalmology
November 2024
Center for Clinical Trials and Evidence Synthesis, Departments of Epidemiology, The Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland, USA; Department of Ophthalmology, The Wilmer Eye Institute, The Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Purpose: Evaluation of longer-term effectiveness of three intravitreal therapies (methotrexate, ranibizumab, or dexamethasone implant) for participants enrolled in the randomized comparative effectiveness trial the Macular Edema Ranibizumab versus Intravitreal anti-inflammatory Therapy (MERIT) Trial followed for24 weeks.
Design: Multicenter randomized controlled clinical trial with masked evaluation of retinal thickness and visual acuity.
Participants: Patients with persistent or recurrent uveitic macular edema.
Rom J Ophthalmol
October 2024
Department of Ophthalmology, Government Medical College and Hospital, Chandigarh, India.
Aim: To evaluate the appropriateness and readability of the medical knowledge provided by ChatGPT-3.5 and Google Bard, artificial-intelligence-powered conversational search engines, regarding surgical treatment for glaucoma.
Methods: In this retrospective, cross-sectional study, 25 common questions related to the surgical management of glaucoma were asked on ChatGPT-3.
Vet World
August 2024
Postgraduate Program in Animal Science, UNOESTE, Presidente Prudente, São Paulo, Brazil.
Background And Aim: Measuring intraocular pressure (IOP) is crucial for identifying potentially damaging changes in the eyes, including diseases as glaucoma and uveitis. This study compared intraocular pressure (IOP) measurements in cats using the Tonovet and Tonovet Plus (rebound), Tono-Pen Avia Vet (applanation), and Kowa HA-2 (Goldman's methodology applanation) tonometers.
Materials And Methods: 55 healthy cats (108 eyes) were assessed through three distinct studies: An experiment (10 eyes of five cats) to correlate IOP manometry and tonometry values and ascertain the correlation coefficient (r2); an study (10 eyes of five sedated cats) to contrast manometer and tonometer readings; and an outpatient clinical trial (80 eyes of 45 cats) to analyze only tonometer measurements.
J Ophthalmic Inflamm Infect
September 2024
Retina Associates of Kentucky, Lexington, KY, USA.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!