AI Article Synopsis

  • A new multielectrode contact-mapping and ablation device, called the Globe, was evaluated for its effectiveness in catheter ablation for atrial fibrillation (AF), combining features from single-tip and balloon catheters.
  • The Globe array consists of 122 electrodes that can simultaneously record data and perform ablation on multiple sites, achieving a high success rate of 99.1% in pulmonary vein isolation (PVI) among 60 patients.
  • Despite a couple of minor complications, the Globe catheter showed promising results, with a 12-month PVI success rate of 75.5% in a subset of patients, suggesting potential for advanced mapping and ablation techniques in the future.

Article Abstract

Introduction: The critical question for technological advancement of catheter ablation of atrial fibrillation (AF) is whether a creative new concept can combine and even improve the options of single-tip catheters with the simplicity of the use of balloon catheters. Herein are described the results from the first clinical study of a new multielectrode contact-mapping plus ablation array (Globe) offering such a complete solution.

Methods And Results: The multielectrode Globe array consists of 16 flat ribs with 122 gold-plated electrodes. Each electrode can record electrograms, ablate, pace, and can measure tissue contact and temperature. Single-shot pulmonary vein isolation (PVI) is possible with temperature-guided ablation of up to 24 electrodes simultaneously with automatic, individual power control of every electrode. Sixty patients with symptomatic AF underwent PVI using the Globe. In all sixty patients, acute PVI was achieved in 232 of 234 attempted PVs (99.1%). In 34 patients treated with "single-hot-shot" ablation, PVI was achieved in 136 of 136 PVs (100%). Single-procedure 12-month freedom from AF off antiarrhythmic drugs in the "single-hot-shot" group was 75.5% and freedom from AF/atrial tachycardia 72.3%. In two patients, pericardial tamponade was observed, one after a transseptal puncture, and one during array insertion with an over-advanced sheath. There were no other device-related serious adverse events, including stroke, PV stenosis, esophageal perforation, or phrenic nerve palsy.

Conclusions: In this first clinical series, the Globe catheter was found to be an easy-to-use system for single-shot PVI. The continuously updated multielectrode voltage and activation mapping data indicate future options for mapping and ablation beyond PVI.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.14172DOI Listing

Publication Analysis

Top Keywords

multielectrode contact-mapping
8
contact-mapping ablation
8
atrial fibrillation
8
sixty patients
8
pvi achieved
8
ablation pvi
8
ablation
6
pvi
6
patients
5
global multielectrode
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!