This article deals with the long-term preclinical validation of WIMAGINE (Wireless Implantable Multi-channel Acquisition system for Generic Interface with Neurons), a 64-channel wireless implantable recorder that measures the electrical activity at the cortical surface (electrocorticography, ECoG). The WIMAGINE implant was designed for chronic wireless neuronal signal acquisition, to be used e.g., as an intracranial Brain-Computer Interface (BCI) for severely motor-impaired patients. Due to the size and shape of WIMAGINE, sheep appeared to be the best animal model on which to carry out long-term validation. The devices were implanted in two sheep for a follow-up period of 10 months, including idle state cortical recordings and Somato-Sensory Evoked Potential (SSEP) sessions. ECoG and SSEP demonstrated relatively stable behavior during the 10-month observation period. Information recorded from the SensoriMotor Cortex (SMC) showed an SSEP phase reversal, indicating the cortical site of the sensorimotor activity was retained after 10 months of contact. Based on weekly recordings of raw ECoG signals, the effective bandwidth was in the range of 230 Hz for both animals and remarkably stable over time, meaning preservation of the high frequency bands valuable for decoding of the brain activity using BCIs. The power spectral density (in dB/Hz), on a log scale, was of the order of 2.2, -4.5 and -18 for the frequency bands (10-40), (40-100), and (100-200) Hz, respectively. The outcome of this preclinical work is the first long-term validation of the WIMAGINE implant, highlighting its ability to record the brain electrical activity through the dura mater and to send wireless digitized data to the external base station. Apart from local adhesion of the dura to the skull, the neurosurgeon did not face any difficulty in the implantation of the WIMAGINE device and post-mortem analysis of the brain revealed no side effect related to the implantation. We also report on the reliability of the system; including the implantable device, the antennas module and the external base station.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6712079PMC
http://dx.doi.org/10.3389/fnins.2019.00847DOI Listing

Publication Analysis

Top Keywords

implantation wimagine
8
wimagine wireless
8
validation wimagine
8
wireless implantable
8
electrical activity
8
wimagine implant
8
long-term validation
8
frequency bands
8
external base
8
base station
8

Similar Publications

Insights into the mechanical interaction between an active cranial implant and the skull subjected to moderate impact loadings.

J Mech Behav Biomed Mater

February 2022

Arts et Métiers Institute of Technology, Université Sorbonne Paris Nord, IBHGC - Institut de Biomécanique Humaine Georges Charpak, HESAM Université, F-75013, Paris, France.

In the context of cochlear implants, which are now widely used, and innovative active devices, the cranial implantation of electronic devices raises new questions about the mechanical interactions between the implant and the skull. The aim of this study was to build a methodology using experimental data and numerical simulations to evaluate the mechanical interactions between the skull and the WIMAGINE® active cranial implant intended for use for tetraplegic patients. A finite element model of the implant housing and a simplified model of the three-layered skull were developed.

View Article and Find Full Text PDF

The evaluation of the long-term stability of ElectroCorticoGram (ECoG) signals is an important scientific question as new implantable recording devices can be used for medical purposes such as Brain-Computer Interface (BCI) or brain monitoring.The long-term clinical validation of wireless implantable multi-channel acquisition system for generic interface with neurons (WIMAGINE), a wireless 64-channel epidural ECoG recorder was investigated. The WIMAGINE device was implanted in two quadriplegic patients within the context of a BCI protocol.

View Article and Find Full Text PDF

Background: To simulate mechanical shocks on an intracranial implant called WIMAGINE, Clinatec chose a Johnson-Cook model to account for the viscoplastic behavior of grade 2 titanium in a dynamic study using Radioss.

Methods: Thirty tensile specimens were subjected to tensile tests at room temperature, and the influence of the strain rate (8 × 10 and 8 × 10 s) and sandblasting was analyzed. Relaxations were included in the tests to analyze viscosity phenomena.

View Article and Find Full Text PDF

Brain-computer interfaces (BCIs) provide a way for the brain to interface directly with a computer. Many different brain signals can be used to control a device, varying in ease of recording, reliability, stability, temporal and spatial resolution, and noise. Electrocorticography (ECoG) electrodes provide a highly reliable signal from the human brain surface, and these signals have been used to decode movements, vision, and speech.

View Article and Find Full Text PDF

This article deals with the long-term preclinical validation of WIMAGINE (Wireless Implantable Multi-channel Acquisition system for Generic Interface with Neurons), a 64-channel wireless implantable recorder that measures the electrical activity at the cortical surface (electrocorticography, ECoG). The WIMAGINE implant was designed for chronic wireless neuronal signal acquisition, to be used e.g.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!