Protective Efficacy of Different Ocular Radiation Protection Devices: A Phantom Study.

Cardiovasc Intervent Radiol

Clinic of Diagnostic and Interventional Radiology, Philipps-University of Marburg, Baldingerstrasse 1, 35043, Marburg, Germany.

Published: January 2020

Purpose: The aim of this study was to investigate the efficacy of different designs and types of ocular radiation protection devices depending on simulated varied body heights in a phantom-simulated thoracic intervention.

Materials And Methods: A clinical angiography system with a standardized fluoroscopy protocol with an anthropomorphic chest phantom as a scattering object and optically stimulated luminescence dosimeters for measuring radiation dose were used. The dosimeters were placed at the position of eyes of an anthropomorphic head phantom simulating the examiner. The head phantom was placed on a height-adjustable stand simulating the height of the examiner from 160 to 200 cm with 10 cm increments. The dose values were then measured with no radiation protection, a weightless-like radiation protection garment, radiation protection glasses and visors.

Results: The average dose reduction using radiation protection devices varied between 57.7 and 83.4% (p < 0.05) in comparison with no radiation protection. Some radiation protection glasses and visors showed a significant dose reduction for the eye lenses when the height of the examiner increased. The right eye was partially less protected, especially if the distances between the simulated examiner's head and the scatter object were small.

Conclusion: All the investigated protection devices showed a significant reduction in radiation exposure to the simulated examiner. For some devices, the radiation dose increased with decreasing distance to the scattering object, especially for the right eye lens.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00270-019-02319-1DOI Listing

Publication Analysis

Top Keywords

radiation protection
24
protection devices
12
ocular radiation
8
head phantom
8
radiation
7
protection
6
protective efficacy
4
efficacy ocular
4
phantom
4
devices phantom
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!