Objectives: This study sought to minimize the risk of permanent pacemaker implantation (PPMI) with contemporary repositionable self-expanding transcatheter aortic valve replacement (TAVR).
Background: Self-expanding TAVR traditionally carries a high risk of PPMI. Limited data exist on the use of the repositionable devices to minimize this risk.
Methods: At NYU Langone Health, 248 consecutive patients with severe aortic stenosis underwent TAVR under conscious sedation with repositionable self-expanding TAVR with a standard approach to device implantation. A detailed analysis of multiple factors contributing to PPMI was performed; this was used to generate an anatomically guided MInimizing Depth According to the membranous Septum (MIDAS) approach to device implantation, aiming for pre-release depth in relation to the noncoronary cusp of less than the length of the membranous septum (MS).
Results: Right bundle branch block, MS length, largest device size (Evolut 34 XL; Medtronic, Minneapolis, Minnesota), and implant depth > MS length predicted PPMI. On multivariate analysis, only implant depth > MS length (odds ratio: 8.04; 95% confidence interval: 2.58 to 25.04; p < 0.001) and Evolut 34 XL (odds ratio: 4.96; 95% confidence interval: 1.68 to 14.63; p = 0.004) were independent predictors of PPMI. The MIDAS approach was applied prospectively to a consecutive series of 100 patients, with operators aiming to position the device at a depth of < MS length whenever possible; this reduced the new PPMI rate from 9.7% (24 of 248) in the standard cohort to 3.0% (p = 0.035), and the rate of new left bundle branch block from 25.8% to 9% (p < 0.001).
Conclusions: Using a patient-specific MIDAS approach to device implantation, repositionable self-expanding TAVR achieved very low and predictable rates of PPMI which are significantly lower than previously reported with self-expanding TAVR.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcin.2019.05.056 | DOI Listing |
Sci Rep
July 2024
Department of Cardiology, Second Affiliated Hospital Zhejiang University School of Medicine, Hangzhou, 310009, People's Republic of China.
Repositionable self-expanding valves allow for repositioning during deployment to achieve optimal valve placement. However, the risk of brain injury associated with repositioning, as detected by diffusion-weighted magnetic resonance imaging (DW-MRI), is unknown. Consecutive patients undergoing transcatheter aortic valve replacement (TAVR) with repositionable self-expanding valves and receiving DW-MRI before and within 7 days post-TAVR procedure were included.
View Article and Find Full Text PDFFront Cardiovasc Med
December 2023
Department of Cardiology, Angiology, Pneumology and Internal Intensive Care, University of Ulm, Ulm, Germany.
Background: Conduction disturbances demanding permanent pacemaker implantation (PPI) remain a common complication after transcatheter aortic valve replacement (TAVR). Optimization of the implantation depth (ID) by introducing the cusp-overlap projection (COP) technique led to a reduced rate of PPI when self-expanding valves were used.
Objectives: The aim of the present study was to determine if using the novel COP view is applicable for all types of TAVR prosthesis and results in a higher ID and reduced incidence of new conduction disturbances and PPI.
Life (Basel)
August 2023
1st Cardiology Department, AHEPA University General Hospital, 54636 Thessaloniki, Greece.
Introduction: The self-expanding, resheathable, repositionable transcatheter aortic heart valve Portico is being used successfully for transcatheter aortic valve implantation procedures (TAVI) in patients with severe aortic stenosis. The aim of this study was to evaluate outcomes at 2 years after TAVI with the Portico valve.
Methods: Multicenter registry of clinical, echocardiographic and survival data from consecutive patients treated with the Portico TAVI system (Abbott, Chicago, IL, USA) in three cath labs in Northern Greece and Epirus during 2017-2020.
JAMA Netw Open
October 2022
Department of Cardiovascular Surgery, Houston Methodist DeBakey Heart and Vascular Center, Houston, Texas.
Importance: Long-term follow-up after transcatheter aortic valve replacement (TAVR) is of interest given that longitudinal data on mortality and durability of transcatheter heart valves are limited. The REPRISE III (Repositionable Percutaneous Replacement of Stenotic Aortic Valve Through Implantation of Lotus Valve System-Randomized Clinical Evaluation) randomized clinical trial compared the mechanically expanded Lotus valve with the self-expanding CoreValve/EvolutR TAVR platforms.
Objective: To describe the final 5-year outcomes of the REPRISE III trial.
JACC Cardiovasc Interv
January 2022
Department of Cardiology, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark. Electronic address:
Objectives: This study sought to evaluate the 30-day and 1-year safety and performance of the Hydra transcatheter aortic valve (THV) (in the treatment of symptomatic severe aortic stenosis in patients at high or extreme surgical risk.
Background: The Hydra THV is a novel repositionable self-expanding system with supra-annular bovine pericardial leaflets.
Methods: The Hydra CE study was a premarket, prospective, multicenter, single-arm study conducted across 18 study centers in Europe and Asia-Pacific countries.