We evaluated the efficacy and safety of short-term fully closed-loop insulin delivery using faster versus standard insulin aspart in type 2 diabetes. Fifteen adults with insulin-treated type 2 diabetes underwent 22 hours of closed-loop insulin delivery with either faster or standard insulin aspart in a double-blind randomized crossover design. Basal-bolus regimen was replaced by model predictive control algorithm-directed insulin delivery based on sensor glucose levels. The primary outcome was time with plasma glucose in target range (5.6-10.0 mmol/L) and did not differ between treatments (mean difference [95% CI] 3.3% [-8.2; 1.7], P = 0.17). Mean glucose and glucose variability were comparable, as was time spent below and above target range. Hypoglycaemia (<3.5 mmol/L) occurred once with faster insulin aspart and twice with standard insulin aspart. Mean total insulin dose was higher with faster insulin aspart (mean difference [95% CI] 3.7 U [0.7; 6.8], P = 0.021). No episodes of severe hypoglycaemia or other serious adverse events occurred. In conclusion, short-term fully closed-loop in type 2 diabetes may require higher dose of faster insulin aspart compared with standard insulin aspart to achieve comparable glucose control.