[A Digital Filtering Method for Eliminating DC Offset in Ambulatory ECG Signals].

Zhongguo Yi Liao Qi Xie Za Zhi

Department of Public Foundation, Bengbu Medical College, Bengbu, 233030.

Published: July 2019

The amplifier is easy to saturation because of polarization voltage in ambulatory ECG acquisition. The traditional way is using analog high-pass filter to eliminate, but the output tends to have a residue. If upgrading high-pass filter cutoff frequency, it will lead to low frequency distortion of ECG signals. In this paper, a Savitzky-Golay (SG) smoothing filter has been designed by combining the single edge point and the polynomial order , which can fit the polarization voltage components of ECG signals, filter useful components and get drift-free ECG signals by using the subtraction algorithm. The results of ECG filtering experiment verify the feasibility of the SG smoothing filter, and show the filtered ECG signal without any losses of useful components.

Download full-text PDF

Source
http://dx.doi.org/10.3969/j.issn.1671-7104.2019.04.005DOI Listing

Publication Analysis

Top Keywords

ecg signals
12
ambulatory ecg
8
polarization voltage
8
high-pass filter
8
smoothing filter
8
ecg
7
filter
5
digital filtering
4
filtering method
4
method eliminating
4

Similar Publications