Background: The identification and precise removal of bony impingement lesions during arthroscopic débridement arthroplasty for elbow osteoarthritis is technically difficult. Surgical navigation systems, combined with preoperative 3-dimensional (3D) assessment of bony impingements, can provide real-time tracking of the surgical instruments and impingement lesions. This study aims to determine the registration accuracy of the navigation system for the humerus and ulna during elbow arthroscopy.
Methods: We tested the registration procedure using resin bone models of 3 actual patients with elbow osteoarthritis. We digitized bone surface points using navigation pointers under arthroscopy. We initially performed paired-point registration, digitizing 6 preset anatomical landmarks, and then refined the initial alignment with surface matching registration, digitizing 30 points. The registration accuracy for each trial was evaluated as the mean target registration error in each reference marker. Three observers repeated the registration procedure 5 times each with the 3 specimens (total, 45 trials). The median of the registration accuracy was evaluated in total (45 trials) as the accuracy of the registration procedure. The differences in the registration accuracy among the 3 observers (median of 15 trials) were also examined.
Results: The total registration accuracies were 0.96 mm for the humerus and 0.85 mm for the ulna. No significant differences were found in the registration accuracy for the humerus and ulna among the 3 observers.
Conclusions: This arthroscopic-assisted registration procedure is sufficiently feasible and accurate for application of the navigation system to arthroscopic débridement arthroplasty in clinical settings.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jse.2019.06.009 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!