Objectives: This study was conducted to investigate the stress around nails and cortical bones in subtrochanteric (ST) fractures fixed using short cephalomedullary nails (CMNs) in finite element models (FEMs) and to determine the appropriate short CMN type for different fracture levels.

Methods: The following three types of short CMNs were used: type A, which is 170 mm in length and has 1 distal locking screw; type B, 200 mm in length and 1 distal screw; and type C, 200 mm in length and 2 distal screws. A total of 24 FEMs were tested on a transverse ST fracture at 8 levels [0, 10, 20, 25, 30, 35, 40 and 50 mm below the lower margin of lesser trochanter (LT)], and were fixed using 3 different CMN types. Finite element analysis was then performed to evaluate the stress around the cortical bones and the CMNs under the assumption of anatomical reduction and fracture gap of 1 mm.

Results: Peak von Mises stress (PVMS) was greatest on the cortical bone around the distal screw hole and was greater than the yield strength at fracture levels ≥ 35 mm below the LT in FEMs fixed with type A and B. In contrast, FEMs fixed with type C showed PVMS less than the yield strength at all fracture levels. The PVMS within the implant was greater than the yield strength at the junction of the nail with the distal screw and distal screw itself at fracture levels ≥ 35 mm below the LT in FEMs fixed using type A. Conversely, in FEMs fixed using type B and C, all PVMSs within the implant were less than the yield strength, regardless of the fracture level.

Conclusion: Short CMNs 170 or 200 mm in length with 1 distal screw may be used in a limited manner in high ST transverse fractures under the assumptions of anatomical reduction and fracture gap ≤ 1 mm. Meanwhile, short CMN 200 mm in length with 2 distal screws may be an available treatment option in most of ST transverse fractures regardless of the fracture level under the same set of assumptions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2019.08.014DOI Listing

Publication Analysis

Top Keywords

fracture levels
20
length distal
20
distal screw
20
200 mm length
16
yield strength
16
fems fixed
16
fixed type
16
finite element
12
strength fracture
12
fracture
11

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!