Background: Advanced bipolar vessel sealing devices are widely used in human and veterinary medicine to reduce the operation time and intraoperative blood loss. Because most devices are made for a single use, their application is cost intensive. The aim of this study was to investigate the influence of multiple uses and sterilization on the performance of bipolar vessel sealing devices.

Methods: The burst pressure of sealed porcine renal arteries was compared between the disposable devices "LigaSure"(Valleylab) and "Caiman"(Braun Vetcare) and the reuseable device "MarSeal"(KLS Martin). Additionally, the influence of the sterilization process was investigated, and the sealing time, number of cutting activations and tissue sticking were noted.

Results: The disposable devices showed reliable performances over multiple activations and sterilization cycles, except for one Caiman device. Seals created with all devices achieved supra-physiologic burst pressures, with the highest pressure measured in a cycle using a MarSeal device.

Discussion: During 25 activations with and without intermittent sterilization, no correlation was found between increasing seal numbers and decreasing burst pressure. However, the number of cycles in our study was limited to five.

Conclusion: For limited numbers of cycles and sterilization procedures, seals created with the disposable vessel sealing devices LigaSure and Caiman achieved burst pressures comparaable to those produced with the reuseable MarSeal.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6701801PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0221488PLOS

Publication Analysis

Top Keywords

vessel sealing
16
bipolar vessel
12
sealing devices
12
multiple sterilization
8
performance bipolar
8
burst pressure
8
disposable devices
8
seals created
8
burst pressures
8
devices
7

Similar Publications

Right coronary sinus of Valsalva pseudoaneurysm after a motorcycle accident: A case report.

Medicine (Baltimore)

January 2025

Department of Radiology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.

Rationale: Traumatic pseudoaneurysm of the sinus of Valsalva (PSV) is a rare but life-threatening condition, often resulting from blunt chest trauma. Rapid progress and a high risk of rupture highlight the importance of prompt diagnosis and intervention. We present a case of a rare pseudoaneurysm linked to the right coronary sinus after blunt chest trauma.

View Article and Find Full Text PDF

Purpose: To report a case series on using a novel semi-branch feature in custom-made stent-grafts in the endovascular treatment of complex aortic aneurysms and summarize the contemporary usage of this technology.

Case Series: Four patients underwent endovascular aortic aneurysm repair (EVAR) with a custom-made semi-branch stent-graft (Semi-Branch Endovascular Aortic Aneurysm Repair [SBEVAR]). Two male patients, 75- and 76-year-old, were treated due to failed EVAR with late-type Ia endoleak, and the other two, 80- and 55-year-old male patients, due to a juxta-renal aortic abdominal aneurysm (JRAAA).

View Article and Find Full Text PDF

Rotablator-associated coronary perforation can be fatal if bailout is delayed. Successful bailout is typically defined as the disappearance of contrast extravasation after a haemostatic intervention. We report a case of recurrent cardiac tamponade in the subacute phase, wherein haemostasis appeared to have been achieved on angiography following the implantation of a covered stent during the index procedure.

View Article and Find Full Text PDF

Aneurysm dome and vessel pressure measurements with coiling, stent assisted coiling and flow diversion.

Acta Neurochir (Wien)

January 2025

Division of Neurosurgery, Beth Israel Deaconess Medical Center, Harvard Medical School, 110 Francis Street , Boston, MA, 02215, USA.

Background: Variability in long-term endovascular treatment outcomes for intracranial aneurysms has prompted questions regarding the effects of these treatments on aneurysm hemodynamics. Endovascular techniques disrupt aneurysmal blood flow and shear, but their influence on intra-aneurysmal pressure remains unclear. A better understanding of aneurysm pressure effects may aid in predicting outcomes and guiding treatment decisions.

View Article and Find Full Text PDF

Iatrogenic intramural esophageal hematoma during EUS-B-FNA procedure.

BMC Pulm Med

January 2025

Unidade de Broncologia e Pneumologia de Intervenção - Instituto Português de Oncologia Francisco Gentil, Coimbra, Portugal.

Background: Esophageal ultrasound with bronchoscope fine needle aspiration (EUS-B-FNA) is a valuable tool for the diagnosis and staging of lung cancer, complementing endobronchial lung ultrasound (EBUS). While generally considered safe, there is a notable lack of comprehensive knowledge within the interventional pulmonology community regarding potential complications.

Case Presentation: We present a case involving a 66-year-old male with squamous cell lung carcinoma undergoing mediastinal staging.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!