Objectives: To compare the performance and time-to-result (TTR) for antimicrobial susceptibility testing (AST) of positive blood cultures (PBC) using the Accelerate Pheno™ system (AXDX) and both a direct VITEK® 2 card inoculation workflow (DV2) and traditional FDA-approved VITEK® 2 workflow using subcultured isolates (V2).
Methods: Patient samples with monomicrobial Gram-negative rod bacteremia were tested on AXDX and DV2 in tandem and compared to V2 AST results. Categorical agreement (CA) errors were adjudicated using broth microdilution. Instrumentation times and AST TTR were compared.
Results: AXDX and DV2 had a CA of 93.4% and 97.4%, respectively, compared to V2. Postadjudication, AXDX, DV2, and V2 had CA of 94.7%, 95.7%, and 96.5%, respectively. Instrument run times were 6.6 h, 9.4 h, and 9.2 h, and AST TTR were 8.9 h, 12.9 h and 35.5 h, respectively.
Conclusions: AXDX and DV2 ASTs are fast and reliable, which may have significant antimicrobial stewardship implications.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.diagmicrobio.2019.05.013 | DOI Listing |
J Assoc Med Microbiol Infect Dis Can
October 2020
Department of Medicine, University of Calgary, Calgary, Alberta, Canada.
Background: Our laboratory uses matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI) and the VITEK 2 system (DV2) directly from positive blood cultures (BC) for organism identification (ID) and antimicrobial susceptibility testing (AST). Our objective was to compare direct MALDI-DV2 with a commercial BC ID-AST platform, the Accelerate Pheno system (AXDX), in the ID-AST of clinical and seeded BC positive for gram-negative bacilli (GNB).
Methods: BC positive for GNB were collected over a 3-mo period and tested using AXDX and direct MALDI-DV2 and compared with conventional methods.
Diagn Microbiol Infect Dis
November 2019
Department of Pathology and Laboratory Medicine, Indiana University School of Medicine, Indianapolis, Indiana, USA.
Objectives: To compare the performance and time-to-result (TTR) for antimicrobial susceptibility testing (AST) of positive blood cultures (PBC) using the Accelerate Pheno™ system (AXDX) and both a direct VITEK® 2 card inoculation workflow (DV2) and traditional FDA-approved VITEK® 2 workflow using subcultured isolates (V2).
Methods: Patient samples with monomicrobial Gram-negative rod bacteremia were tested on AXDX and DV2 in tandem and compared to V2 AST results. Categorical agreement (CA) errors were adjudicated using broth microdilution.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!