Establishment of the intelligent verification criteria for a routine urinalysis analyzer in a multi-center study.

Clin Chem Lab Med

Department of Clinical Laboratory, State Key Laboratory of Molecular Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, No. 17 Panjiayuan Nanli, Chaoyang District, Beijing 100021, P.R. China.

Published: November 2019

Background Although laboratory information system (LIS) is widely used nowadays, the results of routine urinalysis still need 100% manual verification. We established intelligent verification criteria to perform the automated verification process and reduce manual labor. Methods A total of 4610 urine specimens were obtained from the patients of three hospitals in Beijing, China. Firstly, 895 specimens were measured to establish the reference intervals of formed-element parameters in UF5000. Secondly, 2803 specimens were analyzed for setting up the intelligent verification criteria (including the microscopic review rules and manual verification rules). Lastly, 912 specimens were used to verify the efficacy and accuracy of the intelligent verification criteria. Phase-contrast microscopes were used for the microscopic review. Results Employing a results level corresponding relationship in specific parameters including hemoglobin (red blood cell [RBC]), leukocyte esterase (white blood cell [WBC]) and protein (cast) between the dry-chemistry analysis and formed-element analysis, as well as instrument flags, we established seven WBC verification rules, eight RBC verification rules and four cast verification rules. Based on the microscopy results, through analyzing the pre-set rules mentioned earlier, we finally determined seven microscopic review rules, nine manual verification rules and three auto-verification rules. The microscopic review rate was 21.98% (616/2803), the false-negative rate was 4.32% (121/2803), the total manual verification rate was 35.71% (1001/2803) and the auto-verification rate was 64.29% (1802/2803). The validation results were consistent. Conclusions The intelligent verification criteria for urinary dry-chemistry and urinary formed-element analysis can improve the efficiency of the results verification process and ensure the reliability of the test results.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2019-0344DOI Listing

Publication Analysis

Top Keywords

intelligent verification
20
verification criteria
20
verification rules
20
manual verification
16
microscopic review
16
verification
14
rules
9
routine urinalysis
8
verification process
8
review rules
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!