Background: The B-Laser™ atherectomy system (Eximo Medical, Israel) is a 355 nm solid-state Nd:YAG short pulse laser for de-novo and restenotic infrainguinal PAD with enhanced affinity for atheroma and calcified plaque.

Methods: The study was a prospective, single-arm, multi-center, international, open-label study assessing the B-Laser™ in symptomatic (Rutherford 2 to 4) infrainguinal peripheral artery disease. Primary core lab efficacy was mean reduction in diameter stenosis >20% by the B-Laser™ catheter alone. Cardiovascular death, major amputation, target lesion revascularization, WIQ, ABI and Rutherford class were obtained at baseline and out to 6 months. Duplex ultrasound patency (PSVR <2.5), was evaluated by Core Lab.

Results: 97 (77 in USA) PAD subjects (51 male, mean 70.5 years [range 46-86]) with 107 lesions were treated with B-Laser™ (average length 5.4 cm [range 1-24], 29.0% infrapopliteal. 77.6% calcification [26.2% severe], 21.5% chronic total occlusions, 20.6% re-stenotic). Average reduction in residual stenosis post B-Laser™ alone was 33.6 ± 14.2%. Baseline and final stenosis (post laser and adjunctive therapy) were 85.7 ± 12.2% and 17.7 ± 11.0%, respectively. Duplex patency was 96.8% at 30-days and 85.6% at 6 months (95.7% 6-month patency with severe calcification), and did not differ between POBA vs. DCB sub-groups. ABI, Rutherford category and WIQ all improved. There was one MAE and three TLRs out of 101 lesions. No procedural distal embolization was noted and there were no major device-related dissections.

Conclusions: Experience with the B-Laser™ atherectomy system in infrainguinal PAD procedures demonstrates a high level of safety and efficacy for denovo and restenotic infrainguinal arterial lesions.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.28435DOI Listing

Publication Analysis

Top Keywords

eximo medical
8
novel laser-based
4
laser-based catheter
4
catheter peripheral
4
peripheral atherectomy
4
atherectomy 6-month
4
6-month eximo
4
b-laser™
4
medical b-laser™
4
b-laser™ ide
4

Similar Publications

The broad spectrum of clinical manifestations caused by peripheral arterial disease [PAD] and the morphologic heterogeneity of associated atherosclerotic lesions present a considerable management challenge. Endovascular interventions are recognized an effective treatment for PAD. Within this revascularization strategy the role of atherectomy debulking modalities continue to evolve.

View Article and Find Full Text PDF

Background: Contact force (CF) changes after onset of radiofrequency (RF) delivery are not well understood and often ascribed to catheter instability.

Objectives: This study sought to characterize CF changes during RF-based pulmonary vein (PV) isolation.

Methods: Catheter-tip parameters including CF for all RF sessions were extracted from a novel catheter-mapping system from 6 patients undergoing PV isolation.

View Article and Find Full Text PDF

Objective: Wound, Ischemia, and foot Infection (WIfI) staging was established to provide objective classification in patients with chronic limb-threatening ischemia (CLTI) and to predict 1-year major amputation risk. Our goal was to validate WIfI staging using data from the Best Endovascular vs Best Surgical Therapy in Patients with CLTI (BEST-CLI) trial.

Methods: Data from the BEST-CLI Trial, a prospective randomized trial comparing surgical revascularization (OPEN) and endovascular revascularization (ENDO), were used to assess the association of WIfI stage on long-term outcomes in an intention-to-treat analysis.

View Article and Find Full Text PDF

Lesion Morphometry of the Pentaspline Pulsed Field Ablation Catheter: Understanding Catheter Pose, Rotation, and Dosing.

Circ Arrhythm Electrophysiol

December 2024

Helmsley Electrophysiology Center, Department of Cardiology, Icahn School of Medicine at Mount Sinai, New York, NY (K.W., M.N., V.Y.R., J.S.K.).

Background: The pentaspline pulsed field ablation catheter achieves pulmonary vein isolation using 8 stacked, pose-specific applications with rotation. The morphology of pose-specific, single or double applications has not been described.

Methods: One or 2 applications were delivered to select veins and discrete atrial and ventricular sites in 9 swine.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!