Article Abstract

Background: Real-time photoacoustic gas monitoring is used for personnel exposure and environmental monitoring, but its accuracy varies when organic solvents such as alcohol contaminate measurements. This is problematic for anesthetic gas measurements in hospitals, because most disinfectants contain alcohol, which could lead to false-high gas concentrations. We investigated the cross-sensitivities of the photoacoustic gas monitor Innova 1412 (AirTech Instruments, LumaSense, Denmark) against alcohols and alcoholic disinfectants while measuring sevoflurane, desflurane and isoflurane in a laboratory and in hospital during surgery.

Methods: 25 mL ethyl alcohol was distributed on a hotplate. An optical filter for isoflurane was used and the gas monitor measured the 'isoflurane' concentration for five minutes with the measuring probe fixed 30 cm above the hotplate. Then, 5 mL isoflurane was added vaporized via an Anesthetic Conserving Device (Sedana Medical, Uppsala, Sweden). After one-hour measurement, 25 mL isopropyl alcohol, N-propanol, and two alcoholic disinfectants were subsequently added, each in combination with 5 mL isoflurane. The same experiment was in turn performed for sevoflurane and desflurane. The practical impact of the cross-sensitivity was investigated on abdominal surgeons who were exposed intraoperatively to sevoflurane. A new approach to overcome the gas monitor's cross-sensitivity is presented.

Results: Cross-sensitivity was observed for all alcohols and its strength characteristic for the tested agent. Simultaneous uses of anesthetic gases and alcohols increased the concentrations and the recovery times significantly, especially while sevoflurane was utilized. Intraoperative measurements revealed mean and maximum sevoflurane concentrations of 0.61 ± 0.26 ppm and 15.27 ± 14.62 ppm. We replaced the cross-sensitivity peaks with the 10th percentile baseline of the anesthetic gas concentration. This reduced mean and maximum concentrations significantly by 37% (p < 0.001) and 86% (p < 0.001), respectively.

Conclusion: Photoacoustic gas monitoring is useful to detect lowest anesthetic gases concentrations, but cross-sensitivity caused one third falsely high measured mean gas concentration. One possibility to eliminate these peaks is the recovery time-based baseline approach. Caution should be taken while measuring sevoflurane, since marked cross-sensitivity peaks are to be expected.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6689173PMC
http://dx.doi.org/10.1186/s12871-019-0822-7DOI Listing

Publication Analysis

Top Keywords

photoacoustic gas
12
anesthetic gas
12
alcoholic disinfectants
12
gas monitoring
8
gas
8
gas monitor
8
sevoflurane desflurane
8
5 ml isoflurane
8
anesthetic
5
cross-sensitivity
5

Similar Publications

Ppb-Level Photoacoustic Detection of Chloroform Using Four-Microphone Array.

Anal Chem

January 2025

International Joint Laboratory for Integrated Circuits Design and Application, Ministry of Education, School of Physics, Zhengzhou University, Zhengzhou 450001, China.

The photoacoustic spectroscopy (PAS) system commonly enhances the efficiency of optical-acoustic-electrical energy conversion by increasing the laser power, optimizing the resonance characteristics of the photoacoustic cell (PAC), and improving the sensitivity of acoustic sensors. However, conventional systems using a single-microphone or a dual-microphone differential setup for point sampling of the photoacoustic signal fail to account for its spatial distribution, leading to a loss of spatial gain. Drawing on microphone array theory derived from sonar technology, this study, for the first time, presents a PAS sensing system based on a four-microphone array, which is applied to detect chloroform gas.

View Article and Find Full Text PDF

A novel balloon-type photoacoustic cell (BTPAC) is proposed to facilitate the detection limitations of acetylene (CH) gas achieving ppb level. Here, an ellipsoidal photoacoustic cavity is employed as the platform for gas-light interaction. By strategically directing the excitation source towards the focal point of the ellipsoidal cavity, ensuring its trajectory traverses the focal point upon each reflection from the interior walls.

View Article and Find Full Text PDF

A compact and portable gas sensor based on quartz-enhanced photoacoustic spectroscopy (QEPAS) for the detection of methane (C1), ethane (C2), and propane (C3) in natural gas (NG)-like mixtures is reported. An interband cascade laser (ICL) emitting at 3367 nm is employed to target absorption features of the three alkanes, and partial least-squares regression analysis is employed to filter out spectral interferences and matrix effects characterizing the examined gas mixtures. Spectra of methane, ethane, and propane mixtures diluted in nitrogen are employed to train and test the regression algorithm, achieving a prediction accuracy of ∼98%, ∼96%, and ∼93% on C1, C2, and C3, respectively.

View Article and Find Full Text PDF

Fiber-Optic Photoacoustic Gas Microprobe Based on Linear Spot-Type Multipass Cell.

Anal Chem

January 2025

School of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, Liaoning 116024, China.

A linear spot-type multipass cell-enhanced fiber-optic photoacoustic gas microprobe is proposed. To further reduce the volume of the gas chamber and enhance the photoacoustic signal, we designed the cross section of the photoacoustic tube as a slit with a height of 10 mm and a width of 1.5 mm.

View Article and Find Full Text PDF

A highly sensitive sulfur dioxide (SO) photoacoustic gas sensor was developed for the sulfur hexafluoride (SF) decomposition detection in electric power systems by using a novel 266 nm low-cost high-power solid-state pulse laser and a high -factor differential photoacoustic cell. The ultraviolet (UV) pulse laser is based on a passive -switching technology with a high output power of 28 mW. The photoacoustic signal was normalized to the laser power to solve the fluctuation of the photoacoustic signal due to the power instability of the UV laser.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!