Objective: The aim of this study was to determine the feasibility and safety of percutaneous axillary artery (AxA) access using vascular closure devices (VCD) for endovascular repair (ER) of thoraco-abdominal aortic aneurysms (TAAA) using fenestrated, branched, and chimney stent grafts.
Methods: Between September 2013 and December 2017, 40 high risk patients (27 men; mean age: 72.7 ± 7.4 years) with TAAA underwent total percutaneous endovascular aortic repair by percutaneous puncture of the third segment of the left AxA, using a 12F sheath for delivery of the bridging visceral stent grafts. Percutaneous closure was performed using ultrasound guidance and two Perclose ProGlide VCDs per puncture site. Endpoints were device success defined as successful haemostasis of the axillary puncture site by VCD, procedure success defined as endovascularly assisted haemostasis of the AxA, major cerebrovascular and peripheral neurological complications, and 30 day mortality.
Results: The median diameter of the AxA in the third segment was 6.5 mm (range 4.4-10.4 mm). The procedure success was 100%. Device success was achieved in 33 patients (82.5%). Device success in patients with AxA diameter >5 mm was 97%. All patients with diameter of the AxA <5 mm developed stenosis or occlusions of the vessel (n = 6), which were treated by percutaneous stent implantation. One bleeding at the AxA puncture site was treated with trans-femoral implantation of a covered stent. No secondary open procedures were required. No neurological deficit of the left upper limb was seen at discharge. Overall mortality at 30 days was 10%.
Conclusions: Direct puncture of the AxA in the third segment and its use as an access vessel for large sheaths during total percutaneous treatment of complex endovascular aortic procedures appear to be feasible and safe in arteries with a diameter >5 mm. Complications can be managed using endovascular techniques.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejvs.2019.01.011 | DOI Listing |
J Endovasc Ther
January 2025
Department of Vascular Surgery, Swiss Aortic Center Bern, Inselspital, University Hospital of Bern, University of Bern, Bern, Switzerland.
Purpose: To perform a systematic review and meta-analysis of the outcomes of Anaconda fenestrated endograft for the treatment of complex abdominal aortic aneurysms (cAAA).
Material And Methods: A systematic search of all the literature reported until May 2024 was performed according to the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) guidelines. The pooled 30-day mortality rate, technical success rate, reintervention rate as well as bridging stent occlusion rate, and corresponding 95% confidence intervals (CIs) were estimated using fixed or random effect methods.
Sensors (Basel)
January 2025
Key Laboratory of Modern Agricultural Equipment, Ministry of Agriculture and Rural Affairs, Nanjing Institute of Agricultural Mechanization, Nanjing 210014, China.
To address several challenges, including low efficiency, significant damage, and high costs, associated with the manual harvesting of , in this study, a machine vision-based intelligent harvesting device was designed according to its agronomic characteristics and morphological features. This device mainly comprised a frame, camera, truss-type robotic arm, flexible manipulator, and control system. The FES-YOLOv5s deep learning target detection model was used to accurately identify and locate .
View Article and Find Full Text PDFSensors (Basel)
January 2025
The Blavatnik School of Computer Science and AI, Tel Aviv University, Tel Aviv 69978, Israel.
This article surveys the literature on miniature radio transmitters designed to track free-ranging wild animals using emitter-localization techniques. The articles covers the topics of power sources used in such transmitters, including miniature batteries and energy harvesting, techniques for generating the transmitted radio-frequency carrier, techniques for creating short radio pulses and more general on-off schedules, modulation in modern wildlife-tracking transmitters, construction, manufacturing, and tuning techniques, and recent trends in this area. The article also describes the recreation of the first successful wildlife-tracking transmitter, a nontrivial invention that had a profound impact on wildlife ecology, and explores its behavior.
View Article and Find Full Text PDFSensors (Basel)
January 2025
Beijing Tsinghua Changgung Hospital, School of Clinical Medicine, Tsinghua University, 168 Litang Road, Changping District, Beijing 102218, China.
(1) Background: To develop a novel capillary refill time measurement system and evaluate its reliability and reproducibility. (2) Methods: Firstly, the utilization of electromagnetic pressure technology facilitates the automatic compression and instantaneous release of the finger. Secondly, the employment of pressure sensing technology and photoelectric volumetric pulse wave analysis technology enables the dynamic monitoring of blood flow in distal tissues.
View Article and Find Full Text PDFJ Clin Med
January 2025
Department of Ophthalmology, University of Bonn, 53127 Bonn, Germany.
: The PreserFlo™ MicroShunt (PFMS) is a bleb-forming device considered to be less invasive than traditional glaucoma surgery such as trabeculectomy. This study evaluates the 1-year success rates as well as safety profile of PFMS in patients having high intraocular pressure (IOP) and/or glaucoma refractory to drop therapy with a history of prior intravitreal dexamethasone therapy. : A total of 16 eyes after PFMS implantation due to elevated IOP after intravitreal dexamethasone implant (DEX-I) administration were included in this retrospective cohort study.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!