Background: Whole-heart magnetic resonance angiography (MRA) requires sophisticated methods accounting for respiratory motion. Our purpose was to evaluate the image quality of compressed sensing-based respiratory motion-resolved three-dimensional (3D) whole-heart MRA compared with self-navigated motion-corrected whole-heart MRA in patients with known thoracic aorta dilation.
Methods: Twenty-five patients were prospectively enrolled in this ethically approved study. Whole-heart 1.5-T MRA was acquired using a prototype 3D radial steady-state free-precession free-breathing sequence. The same data were reconstructed with a one-dimensional motion-correction algorithm (1D-MCA) and an extradimensional golden-angle radial sparse parallel reconstruction (XD-GRASP). Subjective image quality was scored and objective image quality was quantified (signal intensity ratio, SIR; vessel sharpness). Wilcoxon, McNemar, and paired t tests were used.
Results: Subjective image quality was significantly higher using XD-GRASP compared to 1D-MCA (median 4.5, interquartile range 4.5-5.0 versus 4.0 [2.25-4.75]; p < 0.001), as well as signal homogeneity (3.0 [3.0-3.0] versus 2.0 [2.0-3.0]; p = 0.003), and image sharpness (3.0 [2.0-3.0] vs 2.0 [1.25-3.0]; p < 0.001). SIR with the 1D-MCA and XD-GRASP was 6.1 ± 3.9 versus 7.4 ± 2.5, respectively (p < 0.001); while signal homogeneity was 274.2 ± 265.0 versus 199.8 ± 67.2 (p = 0.129). XD-GRASP provided a higher vessel sharpness (45.3 ± 10.7 versus 40.6 ± 101, p = 0.025).
Conclusions: XD-GRASP-based motion-resolved reconstruction of free-breathing 3D whole-heart MRA datasets provides improved image contrast, sharpness, and signal homogeneity and seems to be a promising technique that overcomes some of the limitations of motion correction or respiratory navigator gating.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6667582 | PMC |
http://dx.doi.org/10.1186/s41747-019-0107-4 | DOI Listing |
Radiol Phys Technol
March 2025
Department of Radiology, Mie University School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
This study aims to develop a computerized classification method for significant coronary artery stenosis on whole-heart coronary magnetic resonance angiography (WHCMRA) images using a 3D convolutional neural network (3D-CNN) with attention mechanisms. The dataset included 951 segments from WHCMRA images of 75 patients who underwent both WHCMRA and invasive coronary angiography (ICA). Forty-two segments with significant stenosis (luminal diameter reduction 75%) on ICA were annotated on WHCMRA images by an experienced radiologist, whereas 909 segments without it were annotated at representative sites.
View Article and Find Full Text PDFMagn Reson Med
February 2025
School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Insights Imaging
September 2024
Department of Radiology, West China Hospital of Sichuan University, Chengdu, China.
Magn Reson Imaging
November 2024
Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, United States; Department of Radiology, Emory University School of Medicine, Atlanta, GA, United States. Electronic address:
Background: 5D, free-running imaging resolves sets of 3D whole-heart images in both cardiac and respiratory dimensions. In an application such as coronary imaging when a single, static image is of interest, computationally expensive offline iterative reconstruction is still needed to compute the multiple 3D datasets.
Purpose: Evaluate how the number of physiologic bins included in the reconstruction affects the computational cost and resulting image quality of a single, static volume reconstruction.
PLoS One
February 2024
Department of Diagnostic and Interventional Radiology and Nuclear Medicine, Section of Pediatric Radiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Cardiac MRI is a crucial tool for assessing congenital heart disease (CHD). However, its application remains challenging in young children when performed at 3T. The aim of this retrospective single center study was to compare a non-contrast free-breathing 2D CINE T1-weighted TFE-sequence with compressed sensing (FB 2D CINE CS T1-TFE) with 3D imaging for diagnostic accuracy of CHD, image quality, and vessel diameter measurements in sedated young children.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!