: New methods of measurement require testing to evaluate test validity. This study compares stereoacuity results of the Lang Stereopad to other common clinical stereoacuity assessments in a normal vision population. : A prospective cross-sectional study was conducted comparing the Lang Stereopad stereoacuity results to results from the Lang II, Frisby, and TNO stereo tests. Mean stereoacuity values and their correlation with inter-ocular visual acuity difference were compared for each stereo test. : 98 subjects (mean age of 33.5 years, SD 14.1; 39 males and 59 females) with normal parameters of visual function underwent multiple stereotest assessments. Median stereoacuity values were the lowest (i.e. more detailed stereoacuity) when using the Frisby stereotest (median 40"; 20-170" [minimum-maximum]) and TNO stereotest (median 60"; 15-480"). In comparison, medians were about double at 100" (50-800") for the Lang Stereopad and greater at 200" (200-200") for the Lang II stereotest. There was no correlation for each stereotest with interocular visual acuity differences. : The Lang Stereopad test is easy to administer and it has certain advantages such as no requirement for additional test glasses. It is a useful assessment to add to the clinical armamentarium for binocular assessment of stereopsis. The Lang Stereopad does not agree well with other stereo tests such as the TNO and Frisby but provides a greater stereoacuity test range than the Lang II. The Lang Stereopad now requires testing in a clinical population in which stereoacuity is a pre-requisite part of the assessment.

Download full-text PDF

Source
http://dx.doi.org/10.1080/09273972.2019.1643893DOI Listing

Publication Analysis

Top Keywords

lang stereopad
28
lang
10
stereoacuity
8
stereoacuity lang
8
stereo tests
8
stereoacuity values
8
visual acuity
8
stereotest median
8
stereopad
7
test
5

Similar Publications

: A clinician's choice of stereotest is influenced by the robustness of the measurement, in terms of sensitivity, specificity and test-retest variability. In relation to the latter aspect, there are limited data on the test-retest variability of these new tests and how they compare to the more commonly used stereotests. Therefore, the aim of the study was to determine the test-retest variability of four different measures of stereoacuity (TNO, Frisby, Lang Stereopad and Asteroid (Accurate STEReotest On a mobIle Device)) and to compare the stereoacuity measurements between the tests in an adult population.

View Article and Find Full Text PDF

In this study, infants' visual processing of depth-inducing stimuli was tested using a new method suitable for experimental settings. Stereograms of the Lang-Stereopad were presented in a timed preferential-looking paradigm to determine infants' preference for a stereogram as compared to a stimulus not inducing an impression of depth. A total of 80 infants were tested at 7 months of age; of these, a sub-sample of 41 infants were tested longitudinally at 4 and 7 months to characterize the developmental trajectory of their preference.

View Article and Find Full Text PDF

: New methods of measurement require testing to evaluate test validity. This study compares stereoacuity results of the Lang Stereopad to other common clinical stereoacuity assessments in a normal vision population. : A prospective cross-sectional study was conducted comparing the Lang Stereopad stereoacuity results to results from the Lang II, Frisby, and TNO stereo tests.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!