A limiting factor in advancement of bone marrow based cell therapies is the lack of characterization of cell products delivered to patients. Using an automated hematology analyzer that can be implemented in clinical setting, the composition of bone marrow aspirates (n = 17 patients) and bone marrow concentrates (n = 12 patients) were assessed. ICC estimates were calculated for measuring reliability. Bone marrow aspirates assessment resulted in excellent reliability for determining white blood cells (ICC - 0.96; 95% CI: 0.92-0.99), red blood cells (ICC - 0.9; 95% CI: 0.77-0.96), platelets (ICC - 0.93; 95% CI: 0.85-0.97) composition. Bone marrow concentrate assessment resulted in excellent reliability for determining white blood cells (ICC - 0.97; 95% CI: 0.93-0.99), platelets (ICC - 0.95; 95% CI: 0.89-0.99) and moderate reliability for red blood cells (ICC - 0.66; 95% CI: 0.36-0.87) composition. Modern automated hematology analyzers could assist to better characterize the cell therapy products to provide reliable and consistent outcomes.