MR safety assessment of active implantable medical devices.

Radiologe

IADI, U1254, INSERM, Université de Lorraine, CHRU de Nancy Brabois, Bâtiment Recherche (anciennement EFS), Rez-de-Chaussé, Rue du Morvan, 54511, Nancy, France.

Published: December 2019

Background: Increasing numbers of patients with active implantable medical devices (AIMDs) require magnetic resonance (MR) examinations. The manufacturers are continuing to improve the MR compatibility of their AIMDs. To this end, a variety of measurement methods and numerical simulations are used to evaluate the risks associated with magnetic resonance imaging (MRI).

Objective: In this article, test methods used to investigate interactions between AIMDs with radio frequency fields and time-varying magnetic gradient fields are reviewed.

Materials And Methods: A literature review of known test methods for radio frequency and gradient field exposure of AIMDs with leads, in particular for neurostimulators, cochlear implants, and implanted infusion pumps, is presented. The state of the art and promising methods are discussed.

Results: ISO/TS 10974 describes the design of high frequency and gradient injection setups to test conductive materials. A large number of sensor designs have been published to measure the induced voltages and currents through radio frequency and gradient fields and for monitoring AIMDs during MR examinations in in vitro tests.

Conclusion: The test methods should be planned to be as conservative as possible to cover the worst case scenario. However, in vitro measurements and computer simulation are far from being able to cover all possible configurations in their complexity and uniqueness. For safer MR examinations, current research recommends in vivo testing prior to MR, parallel radiofrequency transmission techniques, and new sequences with reduced energy input in the presence of AIMDs.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00117-019-0541-6DOI Listing

Publication Analysis

Top Keywords

test methods
12
radio frequency
12
frequency gradient
12
active implantable
8
implantable medical
8
medical devices
8
magnetic resonance
8
gradient fields
8
aimds
6
methods
6

Similar Publications

Determination of antimicrobial susceptibility and virulence-related genes of Trueperella pyogenes strains isolated from various clinical specimens in animals.

Pol J Vet Sci

June 2024

Department of Surgery, Faculty of Veterinary Medicine, University of Siirt, Kezer Campus, Veysel Karani District, University Street, Siirt/Türkiye.

In this study, a total of 32 Trueperella pyogenes strains isolated from different disease specimens in cattle, sheep and goats were examined. Antimicrobial susceptibility of the isolates to 10 antimicrobials were determined using the E-test method and MIC values of the antimicrobials were investigated. The genes that play a role in the antimicrobial resistance or virulence of T.

View Article and Find Full Text PDF

Health Literacy of General Surgery Patients: A Cross-sectional Study.

J Perianesth Nurs

December 2024

Department of Surgical Nursing, Nursing Faculty, Ege University, Izmir, Turkey.

Purpose: Health literacy is a complex issue that affects the health outcomes of surgical patients. This study aimed to determine the health literacy of general surgery patients.

Design: A descriptive cross-sectional study.

View Article and Find Full Text PDF

Proper management of cattle reproduction has a major impact on the efficiency and profitability of dairy production. Ultrasound examination and transrectal palpation or the pregnancy-associated glycoprotein (PAG) test are currently the most commonly used methods for pregnancy diagnosis. However, alternative methods to those mentioned above are constantly being sought in order to minimise stress during the examination, the cost of veterinary services and to reduce the rate of errors in pregnancy diagnosis.

View Article and Find Full Text PDF

Constant antigenic changes, new variants and easy transmission of SARS-CoV-2 virus should acquire greater zoonotic attention and need to remain alert. In this retrospective study the aim was to analyze seropositivity to SARS-CoV-2 in dogs by commercial ELISA. The Virus neutralization test (VNT) was modified for the purpose of confirmation of SARS-CoV-2 antibodies in ELISA-positive dog sera.

View Article and Find Full Text PDF

Background: Flavonoids are among the most important compounds found in plants, since laboratory studies have shown them to be a daily requirement in human diets due to their various health benefits. Therefore, this study focused on extracting, purifying, and measuring the antioxidant activity of the flavonoid quercetin, which is widely found in plants and possesses a variety of biological activities, from different plant sources.

Methods: The extraction of quercetin was performed using several methods (chemical, physical, and enzymatic) and several extraction solutions (water, ethanol, and chloroform) from several plants (spinach, dill, Onion Skin, , sumac, digalkhasab chemri, and leelwi chemri).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!