Objectives: Reports ranged from mixed to marginal tubing wear and spallation effects as a complication of roller pumps in cardiopulmonary bypass (CPB). Because the rollers constantly compress part of the tubing, we sought to determine whether circuit materials behave differently under a 3-h simulation of CPB.

Methods: Two different tubing materials (silicone and Tygon) were tested with a customized experimental circuit, designed to allow in vitro simulation of CPB with priming volumes, pressures, revolutions per minute and temperatures equivalent to the clinical scenario. Samples were analysed with optical and field-emission scanning electron microscopy. We collected 200-ml fluid samples at 4 different times: before starting the CPB (T0), when the predicted revolutions per minute corresponded to about 2 min of CPB (T1), at 90 min (T2) and at 180 min (T3). At the end of CPB, we harvested 2 samples of tubing. Lastly, optical investigations and field-emission scanning electron microscopy observations were used for qualitative and quantitative analysis of circulating fragments.

Results: T2 and T3 fluid samples showed more particles than T1 samples. Significant differences in terms of particle numbers were detected: silicone tubing released more fragments per millilitre than Tygon tubing, with both materials releasing particles from 5 to 500 µm. Silicone tubing was associated with a time-dependent increase in small particles released (P = 0.04), whereas this did not apply to large particles or to Tygon tubing. Yet, bootstrap estimates suggested that silicone tubing was associated with the release of more small particles whereas Tygon tubing released more large particles (both P < 0.01). Unlike silicone, Tygon samples taken from the portion of the circuit not subjected to the action of the roller pump did not show any erosion on their surfaces. Samples of both materials taken from the portion subjected to the compression of the roller pump showed signs of significant deterioration.

Conclusions: Silicone showed a worse spallation performance than Tygon, thus appearing less safe for more complex surgery of prolonged duration or for patients with a prior cerebral ischaemic event. Additional risk and cost-effectiveness comparisons to determine the potential benefits of one type of tubing material over the other are warranted to further expand our findings.

Download full-text PDF

Source
http://dx.doi.org/10.1093/icvts/ivz170DOI Listing

Publication Analysis

Top Keywords

silicone tubing
12
tygon tubing
12
tubing
11
tubing materials
8
revolutions minute
8
field-emission scanning
8
scanning electron
8
electron microscopy
8
fluid samples
8
tubing released
8

Similar Publications

Catheter-associated urinary tract infections (CAUTIs), often caused by biofilm-forming Staphylococcus aureus, present significant clinical challenges. Skt35, a dioxopiperidinamide derivative of cinnamic acid, was investigated for its potential antibacterial and antibiofilm activities against S. aureus biofilms.

View Article and Find Full Text PDF

Exploring diverse biomaterials and implants in the ear, nose, and throat by understanding adverse effects and post-usage events. Literature was obtained from Scopus, PubMed, Google Scholar, and Web of Science. A comprehensive analysis was conducted on original research studies, case reports, and case series spanning from December 2010 to May 2022.

View Article and Find Full Text PDF

A 42 year old Afro-Caribbean man underwent Baerveldt Glaucoma Implant (BGI) surgery for silicone oil induced glaucoma. Three months following initial surgery, the 3-0 prolene ripcord suture was removed. Anterior segment OCT demonstrates the position of the intracameral portion of the tube before and after the 3/0 prolene stent suture (PSS) removal.

View Article and Find Full Text PDF

Accessibility of simulated submillimetre vessels for training supermicrosurgeons is limited by cost and access to micro laboratory facilities. Common simulation techniques include in vivo rat mesenteric artery, ex-vivo cryopreserved artery, ex-vivo chicken wing or thigh and synthetic models such as silicone tubing. The lowest cost and most readily accessible of these is the chicken wing model (Hayashi et al.

View Article and Find Full Text PDF

Catheter-associated urinary tract infections (CAUTIs) account for a large proportion of healthcare-associated infections. CAUTIs, caused by colonization of the catheter surface by uropathogens, are challenging to treat, especially when compounded by antibiotic resistance. One prophylactic strategy that could reduce pathogen colonization is bacterial interference, whereby the catheter surface is coated with non-pathogenic bacteria.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!