Background: Open surgical closure and small-bore suture-based preclosure devices have limitations when used for transcatheter aortic valve replacement, percutaneous endovascular abdominal aortic aneurysm repair, or percutaneous thoracic endovascular aortic aneurysm repair. The MANTA vascular closure device is a novel collagen-based technology designed to close large bore arteriotomies created by devices with an outer diameter ranging from 12F to 25F. In this study, we determined the safety and effectiveness of the MANTA vascular closure device.
Methods And Results: A prospective, single arm, multicenter investigation in patients undergoing transcatheter aortic valve replacement, percutaneous endovascular abdominal aortic aneurysm repair, or thoracic endovascular aortic aneurysm repair at 20 sites in North America. The primary outcome was time to hemostasis. The primary safety outcomes were accessed site-related vascular injury or bleeding complications. A total of 341 patients, 78 roll-in, and 263 in the primary analysis cohort, were entered in the study between November 2016 and September 2017. For the primary analysis cohort, transcatheter aortic valve replacement was performed in 210 (79.8%), and percutaneous endovascular abdominal aortic aneurysm repair or thoracic endovascular aortic aneurysm repair was performed in 53 (20.2%). The 14F MANTA was used in 42 cases (16%), and the 18F was used in 221 cases(84%). The mean effective sheath outer diameter was 22F (7.3 mm). The mean time to hemostasis was 65±157 seconds with a median time to hemostasis of 24 seconds. Technical success was achieved in 257 (97.7%) patients, and a single device was deployed in 262 (99.6%) of cases. Valve Academic Research Consortium-2 major vascular complications occurred in 11 (4.2%) cases: 4 received a covered stent (1.5%), 3 had access site bleeding (1.1%), 2 underwent surgical repair (0.8%), and 2 underwent balloon inflation (0.8%).
Conclusions: In a selected population, this study demonstrated that the MANTA percutaneous vascular closure device can safely and effectively close large bore arteriotomies created by current generation transcatheter aortic valve replacement, percutaneous endovascular abdominal aortic aneurysm repair, and thoracic endovascular aortic aneurysm repair devices.
Clinical Trial Registration: URL: https://www.clinicaltrials.gov . Unique identifier: NCT02908880.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1161/CIRCINTERVENTIONS.119.007258 | DOI Listing |
Open Heart
January 2025
Research Department, Weill Cornell Medicine-Qatar, Doha, Qatar
Background: Transcatheter aortic valve replacement (TAVR) is increasingly used for aortic valve replacement instead of surgical aortic valve replacement (sAVR). We aimed to examine the impact of diabetes on 30-day mortality, 30-day readmission and compare outcomes between TAVR and sAVR.
Methods: Data were extracted from the Nationwide Readmissions Database from 2012 to 2017.
Hellenic J Cardiol
January 2025
Department of Cardiac Surgery Research, Lankenau Institute for Medical Research, Main Line Health, Wynnewood, PA, 19096, USA; Department of Cardiac Surgery, Lankenau Heart Institute, Main Line Health, Wynnewood, PA, 19096, USA. Electronic address:
Objective: Thoracic Aortic Aneurysm (TAA) represents an aortic pathology that is caused by the deranged integrity of the three layers of the aortic wall, and is related to severe morbidity and mortality. Consequently, it is crucial to identify the biomarkers implicated in the pathogenesis and biology of TAA. The aim of the current computational study was to assess the differential gene expression profile of the gap junction proteins (GJPs) in patients with TAA in order to identify novel potential biomarkers for the diagnosis and treatment of this disease.
View Article and Find Full Text PDFJ Vasc Surg
January 2025
Department of Surgery, Division of Vascular and Endovascular Surgery, Boston Medical Center, Boston University Chobanian and Avedisian School of Medicine, Boston, MA, USA. Electronic address:
Objective: Infective endocarditis (IE) is associated with significant morbidity and mortality and places patients at risk for subsequent peripheral vascular emboli. Our goals were to analyze the incidence of peripheral emboli and their associated complications and outcomes.
Methods: A retrospective single-center review of all patients with IE from 2013-2021 was performed.
J Vasc Surg
January 2025
Division of Vascular Surgery, University of Washington, Seattle, WA. Electronic address:
Objective: Post-repair surveillance of ruptured abdominal aortic aneurysm (rAAA) is critical for detecting potential complications. Substantial loss to follow-up has been reported in populations undergoing elective endovascular aortic repair (EVAR); however, there is limited data on follow-up rate among patients presenting with rupture. Thus, we investigated follow-up trends and factors influencing retention at a major academic referral center with a wide service area.
View Article and Find Full Text PDFJ Vasc Surg
January 2025
Divisions of Vascular and Endovascular Surgery, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA. Electronic address:
Objectives: Endovascular aneurysm repair (EVAR) for large infrarenal abdominal aortic aneurysms (AAA) has been associated with worse outcomes compared to EVAR for smaller AAAs. Whether these findings apply to complex AAAs (cAAA) remains uncertain.
Methods: We identified all intact complex EVAR (cEVAR) from 2012-2024 in the Vascular Quality Initiative.