Given the current increase in the incidence of coronary artery disease in younger women as well as the high lifetime risk of developing an x-ray-induced malignancy in this population, we aimed at assessing chest radiation in 206 women ≤55 years old undergoing coronary calcium scoring (CACS) by using a Monte Carlo simulation tool. Our data indicate that the simulated radiation dose of the female breast during CACS depends substantially on the starting position of the x-ray tube, with an almost 2 times excess of breast radiation exposure being measured during anterior-posterior tube positioning. Thus, an additional technical feature taking into account the position of the x-ray tube when acquisition is triggered might be an important tool to reduce radiation exposure of the female breast during CACS.