Background: When engaged in visually demanding tasks, anesthesiologists depend on the auditory display of the pulse oximeter (PO) to provide information about patients' oxygen saturation (SpO2). Current auditory displays are not always effective at providing SpO2 information. In this laboratory study, clinician and nonclinician participants identified SpO2 parameters using either a standard auditory display or an auditory display enhanced with additional acoustic properties while performing distractor tasks and in the presence of background noise.
Methods: In a counterbalanced crossover design, specialist or trainee anesthesiologists (n = 25) and nonclinician participants (n = 28) identified SpO2 parameters using standard and enhanced PO auditory displays. Participants performed 2 distractor tasks: (1) arithmetic verification and (2) keyword detection. Simulated background operating room noise played throughout the experiment. Primary outcomes were accuracies to (1) detect transitions to and from an SpO2 target range and (2) identify SpO2 range (target, low, or critical). Secondary outcomes included participants' latency to detect target transitions, accuracy to identify absolute SpO2 values, accuracy and latency of distractor tasks, and subjective judgments about tasks.
Results: Participants were more accurate at detecting target transitions using the enhanced display (87%) than the standard display (57%; odds ratio, 7.3 [95% confidence interval {CI}, 4.4-12.3]; P < .001). Participants were also more accurate at identifying SpO2 range using the enhanced display (86%) than the standard display (76%; odds ratio, 2.7 [95% CI, 1.6-4.6]; P < .001). Secondary outcome analyses indicated that there were no differences in performance between clinicians and nonclinicians for target transition detection accuracy and latency, SpO2 range identification accuracy, or absolute SpO2 value identification.
Conclusions: The enhanced auditory display supports more accurate detection of target transitions and identification of SpO2 range for both clinicians and nonclinicians. Despite their previous experience using PO auditory displays, clinicians in this laboratory study were no more accurate in any SpO2 outcomes than nonclinician participants.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1213/ANE.0000000000004267 | DOI Listing |
J Exp Psychol Hum Percept Perform
January 2025
Department of Psychology, Saarland University.
Task-irrelevant sounds that are semantically congruent with the target can facilitate performance in visual search tasks, resulting in faster search times. In three experiments, we tested the underlying processes of this effect. Participants were presented with auditory primes that were semantically congruent, neutral, or incongruent to the visual search target, and importantly, we varied the set size of the search displays.
View Article and Find Full Text PDFAlzheimers Dement
December 2024
Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIH,, Baltimore, MD, USA.
Background: Cognitive impairment with age remains undetected until it interferes daily life activity or presents dementia symptoms. In the US, 61% of dementia population is not diagnosed, which is in part due to limited sensitivity of clinical neuroimaging modalities in assessing early gray matter (GM) changes. Here we look at microstructural changes in GM using mean apparent propagator (MAP-MRI) in cognitively underperforming (CU) and healthy aging (HA) cohorts, grouped according to their cognitive performance based on the NIH Toolbox.
View Article and Find Full Text PDFAlzheimers Dement
December 2024
Noesis Cognitive Center & Tech Solutions Ltd, Nicosia, Cyprus.
Background A 92-year-old retired seamstress, born in 1932, with 12 years of education, had been residing in a long-term care facility since 2019, following a fall and hip fracture. Post-admission, her cognitive function gradually declined and she did not participate in residential home activities. This study explores the outcomes of an 8-month, multisensory remediation program.
View Article and Find Full Text PDFPsychon Bull Rev
January 2025
Department of Psychology, University of Milano - Bicocca, Piazza dell'Ateneo Nuovo 1, Milan, MI, 20126, Italy.
Auditory iconic words display a phonological profile that imitates their referents' sounds. Traditionally, those words are thought to constitute a minor portion of the auditory lexicon. In this article, we challenge this assumption by assessing the pervasiveness of onomatopoeia in the English auditory vocabulary through a novel data-driven procedure.
View Article and Find Full Text PDFAutism Res
December 2024
Psychiatry and Addictology Department, CIUSSS-NIM Research Center, University of Montreal, Montreal, Quebec, Canada.
Child-directed speech (CDS), which amplifies acoustic and social features of speech during interactions with young children, promotes typical phonetic and language development. In autism, both behavioral and brain data indicate reduced sensitivity to human speech, which predicts absent, decreased, or atypical benefits of exaggerated speech signals such as CDS. This study investigates the impact of exaggerated fundamental frequency (F0) and voice-onset time on the neural processing of speech sounds in 22 Chinese-speaking autistic children aged 2-7 years old with a history of speech delays, compared with 25 typically developing (TD) peers.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!