  • Cerebrospinal fluid (CSF) leakage can cause serious brain and spine issues, highlighting the need for quick diagnostic methods.
  • The study utilizes an immunochromatographic assay (ICA) to detect beta-2 transferrin (β2TF), a specific biomarker for CSF leakage, distinguishing it from serum sialotransferrin (sTF) using a unique sample-preparation step.
  • The ICA showed high accuracy in identifying CSF leakage, with a specificity of 97.1% and a sensitivity of 96.2%, making it a promising point-of-care testing option, although it still takes 70 minutes to complete, which is better than traditional methods.

Article Abstract

Cerebrospinal fluid (CSF) leakage can lead to brain and spine pathologies and there is an urgent need for a rapid diagnostic method for determining CSF leakage. Beta-2 transferrin (β2TF), asialotransferrin, is a specific CSF glycoprotein biomarker used to determine CSF leakage when distinguished from serum sialotransferrin (sTF). : We detected β2TF using an immunochromatographic assay (ICA), which can be potentially developed as a point-of-care (POC) testing platform. Sialic acid-specific lectin selectively captures sTF in multiple deletion lines within an ICA test strip, enabling the detection of β2TF. A sample pre-treatment process efficiently captures excess sTF increasing sensitivity for CSF leakage detection. : An optimal cut-off value for determining the presence of CSF in test samples was obtained from receiver operating characteristic (ROC) analysis of the ratio of the test signal intensity and the deletion lines. On 47 clinical samples, ICA test strips discriminated CSF positive from negative samples with statistically significant (positive negative t-test; =0.00027). Additional artificial positive samples, prepared by mixing CSF positive and negative clinical samples, were used as a further challenge. These positive samples were clearly discriminated from the negative samples (mixture negative t-test; =0.00103) and CSF leakage was determined with 97.1% specificity and 96.2% sensitivity. : ICA represents a promising approach for POC diagnosis of CSF leakage. While requiring 70 min assay time inconvenient for POC testing, our method was significantly shorter than conventional electrophoresis-based detection methods for β2TF.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6592183PMC
http://dx.doi.org/10.7150/thno.34411DOI Listing

Publication Analysis

Top Keywords

csf leakage
24
positive negative
12
csf
10
cerebrospinal fluid
8
immunochromatographic assay
8
poc testing
8
deletion lines
8
ica test
8
clinical samples
8
csf positive
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!