AI Article Synopsis

  • A study was conducted to evaluate the impact of implantable cardioverter defibrillators (ICDs) on patients with continuous flow left ventricular assist devices (LVADs), using data from the Cleveland Clinic between 2004 and 2017.
  • Of the 486 patients analyzed, 79.6% had ICDs prior to LVAD implantation, but the study found no significant associations between the use of ICDs and mortality rates.
  • The findings suggest that while ICDs are associated with various complications and procedures, they do not provide a mortality benefit for patients with continuous flow LVADs.

Article Abstract

Background The effect of implantable cardioverter defibrillators ( ICD ) in patients with continuous flow left ventricular assist devices ( LVAD s) on outcomes has not been evaluated in a randomized clinical trial. Methods and Results This is a retrospective single-center study that included patients who underwent continuous flow LVAD implantation at the Cleveland Clinic between October 2004 and March 2017. Patients were evaluated according to the presence or absence of ICD at the time of LVAD insertion. Among 486 patients in the study cohort, 387 (79.6%) had an ICD before LVAD insertion. Patients with ICD before LVAD were older and had lower use of pre- LVAD inotropes, extracorporeal membrane oxygenation, and mechanical ventilation. There were 81 patients (21.4% of patients with ICD ) who required 93 procedures after LVAD : 74 generator exchanges, 12 lead revisions, and 7 complete system removals because of infection. Of the 99 patients without ICD , 52 (53%) underwent ICD implantation: 29 for primary prevention and 23 for secondary prevention. Patients were followed for a median of 401 (interquartile range 150-966) days. The presence of a pre- LVAD ICD was not associated with mortality in a multivariable model (hazard ratio 1.19, 95% CI 0.73-1.93, P=0.492), nor was the presence of an ICD at any point when analyzed as a time-varying covariate (hazard ratio 1.05, 95% CI 0.50-2.20, P=0.907). Conclusions There is no apparent mortality benefit associated with an ICD in a contemporary cohort of patients with continuous flow LVAD s to balance considerable morbidity involving ICD -related procedures and complications.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6662142PMC
http://dx.doi.org/10.1161/JAHA.118.011813DOI Listing

Publication Analysis

Top Keywords

continuous flow
16
patients continuous
12
patients icd
12
patients
11
icd
11
lvad
9
implantable cardioverter
8
cardioverter defibrillators
8
flow left
8
left ventricular
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!