Background: In a recent study, we determined that 30% of frequent plateletpheresis donors collected using the Trima Accel Automated Blood Collection System (Terumo BCT) had a CD4+ T-cell count below 200 cells/μL. Whether CD4+ T-cell lymphopenia is associated with donation using other plateletpheresis instruments is unknown.

Study Design And Methods: We obtained blood samples from 30 current frequent Fenwal Amicus plateletpheresis donors. All participants had made 20 to 24 plateletpheresis donations in the most recent 365-day period, and all had previously donated over 50 times on the Fenwal Amicus instrument. Blood samples were analyzed to determine blood counts, including CD4+ and CD8+ counts.

Results: Of 30 study participants, none had a CD4+ count below 200 cells/μL. There was one participant with a CD4+ count between 200 and 300 cells/μL. This individual was over the age of 55 and had a history of more than 300 lifetime plateletpheresis sessions. One participant had a CD8+ count below the lower limit of normal (125 cells/μL) and a normal CD4+ count.

Conclusion: We did not detect severe CD4+ lymphopenia in frequent platelet donors undergoing plateletpheresis with the Fenwal Amicus. Since the Fenwal Amicus does not incorporate a leukoreduction system chamber, this finding supports the hypothesis that such chambers-found in the Trima Accel instrument-contribute to CD4+ lymphopenia in frequent platelet donors.

Download full-text PDF

Source
http://dx.doi.org/10.1111/trf.15441DOI Listing

Publication Analysis

Top Keywords

fenwal amicus
20
cd4+ t-cell
12
plateletpheresis donors
12
count 200
12
severe cd4+
8
t-cell lymphopenia
8
frequent plateletpheresis
8
donors collected
8
trima accel
8
cd4+
8

Similar Publications

T-cell lymphopenia in frequent volunteer platelet donors.

Hematology Am Soc Hematol Educ Program

December 2023

Brigham and Women's Hospital, Boston, MA.

In the United States, more than 2 000 000 apheresis platelet units are collected annually from volunteer donors. Platelet donors in the United States and elsewhere are permitted to donate up to 24 times per year. Recently, frequent apheresis platelet donation has been associated with severe T-cell lymphopenia.

View Article and Find Full Text PDF

Background: Severe T-cell lymphopenia of uncertain clinical significance has been observed in frequent apheresis platelet donors. Two commonly used plateletpheresis instruments are the Trima Accel, which uses a leukoreduction system (LRS) chamber to trap leukocytes and the Fenwal Amicus, which does not use an LRS chamber.

Study Design And Methods: We performed an international, multicenter, observational study comparing T-cell populations in frequent platelet donors collected exclusively using the Trima instrument (n = 131) or the Amicus instrument (n = 77).

View Article and Find Full Text PDF

Comparison of hematopoietic progenitor cell collection using different inlet flow rates with the Fenwal Amicus.

J Clin Apher

June 2022

Department of Laboratory Medicine and Pathology, Division of Transfusion Medicine, Mayo Clinic, Rochester, Minnesota, USA.

Purpose: We have used a hematopoietic progenitor cell (HPC) algorithm (standard [STD]) that restricted the inlet flow rate to 65 mL/min for peripheral white blood cell count (PWBC) >35 × 10 /L (STD). In this study, we evaluated a technique that allows 85 mL/min, regardless of the PWBC count (high). For patients with PWBC >35 × 10 /L, a prospective, randomized comparison of the high flow rate vs the STD PWBC-based flow rate (65 mL/min) was performed, comparing CD34+ and lymphocyte yields, collection efficiencies (CE1), mononuclear cells (MNC), and granulocytes, red blood cell (RBC), and platelet content.

View Article and Find Full Text PDF

Purpose: A new software release of the Fenwal Amicus device is now available for red blood cell (RBC) apheresis, offering RBC exchange, RBC depletion (RBCd)/RCB exchange, and RBCd procedures. The main goal of this study was to validate the new RBCd program through the accuracy of the patient's postprocedure hematocrit (HCT) (actual end HCT) to the HCT reported by the device at the end of the procedure (target end HCT). Secondary objectives were to assess the device-related significant adverse events, the patient's cellular losses, and the degree of device induced hemolysis.

View Article and Find Full Text PDF

Background: Collection efficiency (CE) of peripheral blood stem cell (PBSC) collections is negatively affected by increasing white blood cell (WBC) counts of the patient. This study compared a new optimized mononuclear cell (MNC) collection protocol (OPP) to the standard MNC collection protocol recommended by the manufacturer (STP) for PBSC collection in patients with WBC counts >35 000/μL.

Study Design And Methods: Single-center, retrospective, and observational study of 81 autologous PBSC collections on Fenwal Amicus cell separators in 70 adult patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!