NanoBlow: A Simple Device To Limit Contaminants during NanoLC-MS.

J Proteome Res

School of Life and Environmental Sciences, Charles Perkins Centre , University of Sydney, Sydney , NSW 2006 , Australia.

Published: August 2019

Limiting contamination of liquid chromatography/tandem mass spectrometry (LC-MS) systems and reducing the downtime associated with maintenance and cleaning is essential for productivity. We developed a simple device that creates a gas curtain barrier to prevent ions from entering the MS inlet. The gas can be quickly and easily applied when certain contaminant ions are known to elute. We show that the device can prevent the buildup of contaminants on the heated transfer capillary following >100 injections of a crude tissue lysate and improves peptide identifications. The device may provide a promising approach toward improving instrument robustness.

Download full-text PDF

Source
http://dx.doi.org/10.1021/acs.jproteome.9b00175DOI Listing