Objectives: To evaluate the influence of baseline parameters on the occurrence of stent-graft surface movement after endovascular aneurysm repair (EVAR) and to investigate its association with migration and stent-graft-related endoleaks (srEL).

Methods: In this retrospective, cross-sectional study, three-dimensional surface models of the stent-graft, delimited by landmarks using custom-built software, were derived from the pre-discharge and last follow-up computed tomography angiography (CTA). Stent-graft surface movement in the proximal anchoring zone between these examinations was considered significant at a threshold of 9 mm. The Cox proportional hazards model was used to determine baseline variables associated with the occurrence of stent-graft surface movement. The association between migration and srEL with stent-graft surface movement was tested with the chi-square and the Fisher exact test, respectively.

Results: Stent-graft surface movement was observed in 54 (28.9%) of 187 patients. Multivariate analysis revealed that age ([HR] 1.05; p = 0.017), proximal neck diameter ([HR] 5.07; p < 0.001), infrarenal aortic neck angulation ([HR] 1.02, p = 0.002), and proximal neck length ([HR] 0.62, p < 0.001) were significantly associated with the occurrence of stent-graft surface movement. Migration and srEL occurred in 17 (31.5%) and 5 (9.3%) patients, with and 11 (8.3%) and 2 (1.5%) without stent-graft surface movement (p < 0.001, p = 0.022).

Conclusions: Age, neck diameter, infrarenal neck angulation, and proximal neck length were significantly associated with the occurrence of stent-graft surface movement. Apart from possible use of adjunctive sealing systems, concerned patients may benefit from regular CTA surveillance, enabling timely diagnosis of subtle changes of stent-graft position.

Key Points: • Stent-graft surface movement, demonstrating subtle, three-dimensional changes in stent-graft position in the proximal anchoring zone, can be derived from CTA examinations. • Age, proximal neck diameter, and infrarenal neck angulation were significantly associated with an increased incidence of stent-graft surface movement. Stent-graft surface movement is significantly more frequent in patients with stent-graft migration and stent-graft-related endoleaks. • Consideration of risk factors for stent-graft surface movement may help to identify patients who might benefit from regular CTA surveillance and timely diagnosis of subtle changes of stent-graft position, enabling re-interventions to prevent migration and srEL.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6828830PMC
http://dx.doi.org/10.1007/s00330-019-06282-wDOI Listing

Publication Analysis

Top Keywords

stent-graft surface
24
surface movement
24
association migration
12
movement endovascular
8
endovascular aneurysm
8
aneurysm repair
8
baseline parameters
8
migration stent-graft-related
8
stent-graft-related endoleaks
8
occurrence stent-graft
8

Similar Publications

Article Synopsis
  • The study aimed to assess how double chimney thoracic endovascular aortic repair (ch-TEVAR) affects gutter phenomena and endograft deformations in a model that simulates real blood flow conditions.
  • Researchers compared two types of chimney grafts (balloon-expandable and self-expandable) using advanced imaging techniques to evaluate their behavior during different phases of the heart cycle.
  • Results showed that balloon-expandable grafts had significantly larger gutter volumes and exhibited different deformation patterns than self-expandable grafts, highlighting important design considerations for future devices to mitigate risks like endoleak.
View Article and Find Full Text PDF
Article Synopsis
  • The study focuses on the use of transcatheter electrosurgical aortic septotomy (TEAS) to improve endovascular repair outcomes for patients with chronic dissecting aortoiliac aneurysms, addressing challenges like a rigid septum and compressed true lumen.
  • Over 17 patients underwent TEAS from 2021 to 2023, and the main goal was to achieve technical success, evaluated alongside various secondary outcomes related to vessel stability and aneurysm-related complications.
  • Results showed a 100% technical success rate with no major complications, and all target vessels were successfully stented, allowing safe further repairs such as thoracic endovascular aneurysm repair and fenestrated/branched endovascular aneurysm repair.
View Article and Find Full Text PDF

The G-Branch endograft is a novel multibranched "off-the-shelf" device used to repair thoracoabdominal aortic aneurysms (TAAAs). This report describes the hemodynamic and morphological performance of the G-Branch endograft in a human patient with TAAA. We retrospectively reviewed the computed tomography angiography scans and clinical data of a woman in whom TAAA was treated using a G-Branch endograft.

View Article and Find Full Text PDF

Thoracic endovascular aortic repair (TEVAR) is a well-established technique for the management of blunt thoracic aortic injury (BTAI). Despite improvements in vascular imaging, graft material properties, and implant techniques, stent-graft deployment artificially induces aortic stiffening. This study aimed to evaluate the midterm effect of thoracic endovascular aortic repair after blunt thoracic aortic injury on aortic stiffness and cardiac function in young patients using cardiovascular magnetic resonance (CMR) imaging.

View Article and Find Full Text PDF

The ViaHole technique: a novel approach for recanalizing major side branch occluded by Viabahn stent-graft.

CVIR Endovasc

July 2023

Department of Cardiology, Asia Medical Group, Sapporo Heart Center, Sapporo Cardio Vascular Clinic, North 49, East 16, 8-1, Higashi Ward, Sapporo, Hokkaido, 007-0849, Japan.

Introduction: In managing arterial rupture, stent-graft implantation may cause limb ischemia by crossing a major branch for hemostasis. The ViaHole technique could circumvent a major branch occlusion.

Materials And Methods: The process involved advancing retrograde devices into an occluded major branch by the stent-graft implantation to reach the outer surface of the stent-graft, puncturing the stent-graft with a 20-gauge needle to touch the retrograde device, manipulating the guidewire through the needle hole and externalizing it, advancing the microcatheter into the proximal lumen, catching the microcatheter using an antegrade 4-Fr catheter, inserting an antegrade guidewire into the retrograde microcatheter to cross the stent-graft hole, dilating the lesion and stent-graft hole using a 3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!