Nasal sound pressure as objective verification of implant in active transcutaneous bone conduction devices.

Med Devices (Auckl)

Department of Otorhinolaryngology, Head and Neck Surgery, Sahlgrenska University Hospital, The Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.

Published: May 2019

Active transcutaneous bone conduction devices consist of an external audio processor and an internal implant under intact skin. During the surgical procedure, it is important to verify the functionality of the implant before the surgical wound is closed. In a clinical study with the new bone conduction implant (BCI), the functionality of the implant was tested with an electric transmission test, where the output was the nasal sound pressure (NSP) recorded in the ipsilateral nostril. The same measurement was performed in all follow-up visits to monitor the implant's functionality and transmission to bone over time. The objective of this study was to investigate the validity of the NSP method as a tool to objectively verify the implant's performance intraoperatively, as well as to follow-up the implant's performance over time. Thirteen patients with the BCI were included, and the NSP measurement was part of the clinical study protocol. The implant was electrically stimulated with an amplitude-modulated signal generator using a swept sine 0.1-10 kHz. The NSP was measured with a probe tube microphone in the ipsilateral nostril. The NSP during surgery was above the noise floor for most patients within the frequency interval 0.4-5 kHz, showing NSP values for expected normal transmission of a functioning implant. Inter-subject comparison showed large variability, but follow-up results showed only minor variability within each subject. Further investigation showed that the NSP was stable over time. The NSP method is considered applicable to verify the implant's functionality during and after surgery. Such a method is important for implantable devices, but should be simplified and clinically adapted. Large variations between subjects were found, as well as smaller variability in intra-subject comparisons. As the NSP was found to not change significantly over time, stable transmission to bone, and implant functionality, were indicated.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6553998PMC
http://dx.doi.org/10.2147/MDER.S197919DOI Listing

Publication Analysis

Top Keywords

bone conduction
12
nsp
9
nasal sound
8
sound pressure
8
implant
8
active transcutaneous
8
transcutaneous bone
8
conduction devices
8
functionality implant
8
clinical study
8

Similar Publications

Aims And Objectives: This study aimed to investigate the presence, type, and severity of hearing losses in individuals with Duane Retraction Syndrome (DRS), and to ascertain if there are anomalies in the auditory pathways at the brainstem level in DRS, believed to arise from aberrant interaction between cranial nerves and brainstem nuclei.

Study Design: Cross-sectional observational study.

Setting: Tertiary referral centre.

View Article and Find Full Text PDF

Objectives: Cervical vestibular evoked myogenic potentials (cVEMPs) reflect saccular stimulation that results in an inhibitory muscle reflex recorded over the sternocleidomastoid muscle. These responses are utilized to study basic vestibular functions and are also applied clinically. Traditionally, cVEMPs have utilized transient stimuli such as clicks and tonebursts to evoke onset responses.

View Article and Find Full Text PDF

Purpose: This study analyzes temporal bone thickness in pediatric craniofacial anomalies patients to determine the optimal age for preoperative high-resolution computed tomography (HRCT) and enhance surgical planning of bone conduction hearing implants.

Methods: A retrospective examination of 370 temporal bone HRCT scans from 158 pediatric patients with craniofacial anomalies was conducted. Statistical analysis included descriptive statistics and correlation between age and skull thickness.

View Article and Find Full Text PDF

Objectives: Auditory brainstem response (ABR) is the gold standard to assess hearing loss in pediatric patients. Multiple widely accepted ABR protocols with varying parameters are accepted, difference in standards may lead to misdiagnosis or delay in diagnosis and treatment. This study investigates the quality of ABR testing in pediatric patients in addition to changes in diagnoses and management.

View Article and Find Full Text PDF

Objectives: Implantable bone conduction hearing devices offer excellent auditory rehabilitation. Transcutaneous devices, which use an implanted magnet, are gaining popularity due to higher skin complications associated with traditional percutaneous devices. The Cochlear Baha® Attract System (Cochlear Corporation, Sydney, Australia) is a transcutaneous device and is regarded as a passive transcutaneous implant.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!