Dual-Energy CT: Lower Limits of Iodine Detection and Quantification.

Radiology

From the Departments of Imaging Physics (M.C.J., D.D.C., R.R.L.), Interventional Radiology (E.N.K.C., D.L.B.), Diagnostic Radiology, Abdominal Imaging Section (E.P.T.), and Diagnostic Radiology, Neuroradiology Section (D.S.), The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030; and Radiology, Medical Physics Section (X.D.), University of Texas Southwestern Medical Center, Dallas, Tex.

Published: August 2019

Background Assessments of the quantitative limitations among the six commercially available dual-energy (DE) CT acquisition schemes used by major CT manufacturers could aid researchers looking to use iodine quantification as an imaging biomarker. Purpose To determine the limits of detection and quantification of DE CT in phantoms by comparing rapid peak kilovoltage switching, dual-source, split-filter, and dual-layer detector systems in six different scanners. Materials and Methods Seven 50-mL iohexol solutions were used, with concentrations of 0.03-2.0 mg iodine per milliliter. The solutions and water sample were scanned five times each in two phantoms (small, 20-cm diameter; large, 30 × 40-cm diameter) with six DE CT systems and a total of 10 peak kilovoltage settings or combinations. Iodine maps were created, and the mean iodine signal in each sample was recorded. The limit of blank (LOB) was defined as the upper limit of the 95% confidence interval of the water sample. The limit of detection (LOD) was defined as the concentration with a 95% chance of having a signal above the LOB. The limit of quantification (LOQ) was defined as the lowest concentration where the coefficient of variation was less than 20%. Results The LOD range was 0.021-0.26 mg/mL in the small phantom and 0.026-0.55 mg/mL in the large phantom. The LOQ range was 0.07-0.50 mg/mL in the small phantom and 0.20-1.0 mg/mL in the large phantom. The dual-source and rapid peak kilovoltage switching systems had the lowest LODs, and the dual-layer detector systems had the highest LODs. Conclusion The iodine limit of detection using dual-energy CT systems varied with scanner and phantom size, but all systems depicted iodine in the small and large phantoms at or below 0.3 and 0.5 mg/mL, respectively, and enabled quantification at concentrations of 0.5 and 1.0 mg/mL, respectively. © RSNA, 2019 See also the editorial by Hindman in this issue.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6694721PMC
http://dx.doi.org/10.1148/radiol.2019182870DOI Listing

Publication Analysis

Top Keywords

peak kilovoltage
12
detection quantification
8
rapid peak
8
kilovoltage switching
8
dual-layer detector
8
detector systems
8
water sample
8
limit detection
8
mg/ml small
8
small phantom
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!