Purpose: In the practice code of dosimetry, humidity effect is assumed to be constant as far as the measurements are performed in the relative humidity (RH) range of (20-80)%; thus, the humidity effect can be ignored with a dose uncertainty of 0.15%. This assumption is based on the previous experimental results by Niatel and Guiho. Rogers and Ross calculated the stopping power ratio of humid air and dry air for high-energy electron beams by using a Monte Carlo code. They demonstrate that the W value, the mean energy required to create an ion pair in air, is independent of the beam quality when the air is dry, and that the traditional humidity correction can be used also for high-energy photon and electron beams; however, this was only a computational study. In the present study, we measured the humidity correction of Farmer-type ionization chambers in high-energy photon beams and determined the W values of humid air using the calculated energy deposition of humid air with a Monte Carlo code. Furthermore, we proposed an analytical expression to determine a practical humidity correction for an ionization chamber as a function of absolute humidity.

Method: Experiments were carried out using a clinical linear accelerator (linac, Elekta Precise) at the National Metrology Institute of Japan (NMIJ). A shield box was constructed downstream of the linac and connected to an air processor, which maintained the temperature around 22°C and controlled the humidity in the range of (10-70)% inside the box. We prepared two Farmer-type ionization chambers: PTW 30013 and Exradin A19. Each ionization chamber was placed inside the box and irradiated with 6-, 10-, and 15-MV high-energy photon beams from the clinical linac. The energy deposition to the humid air inside the ionization chamber was calculated using the Electron Gamma Shower Version 5 (EGS5) code system.

Results: Stabilization for the humidity of the ionization chamber was completed within 3 h. The polarity and ion recombination corrections did not show any change in the humidity range studied. The measured humidity correction and the evaluated W values of humid air in high-energy photon beams were in good agreement with those by Rogers in TG-21 and by Niatel in the range of RH (10-70)%.

Conclusion: Humidity correction of ionization chambers in high-energy photon beams from the clinical linac was determined experimentally. Using the analytical expression for the energy depositions by EGS5, the analytical expression for the W values was also derived.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13665DOI Listing

Publication Analysis

Top Keywords

high-energy photon
24
photon beams
20
humid air
20
humidity correction
20
ionization chambers
16
ionization chamber
16
humidity
12
humidity range
12
analytical expression
12
air
9

Similar Publications

Cavity as radio telescope for galactic dark photon.

Sci Bull (Beijing)

January 2025

School of Physics and State Key Laboratory of Nuclear Physics and Technology, Peking University, Beijing 100871, China; Center for High Energy Physics, Peking University, Beijing 100871, China; Key Laboratory of Particle Acceleration Physics and Technology, Chinese Academy of Sciences, Beijing 100049, China. Electronic address:

Dark photons, as a minimal extension of the Standard Model through an additional Abelian gauge group, may propagate relativistically across the galaxy, originating from dark matter decay or annihilation, thereby contributing to a galactic dark photon background. The generation of dark photons typically favors certain polarization modes, which are dependent on the interactions between dark matter and dark photons. We introduce a framework in which a resonant cavity is utilized to detect and differentiate these polarizations, leveraging the daily variation in expected signals due to the anisotropic distribution of dark photons and the rotation of the Earth.

View Article and Find Full Text PDF

[Vacuum ultraviolet laser dissociation and proteomic analysis of halogenated peptides].

Se Pu

February 2025

CAS Key Laboratory of Separation Sciences for Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian 116023, China.

Chemical modifications are widely used in research fields such as quantitative proteomics and interaction analyses. Chemical-modification targets can be roughly divided into four categories, including those that integrate isotope labels for quantification purposes, probe the structures of proteins through covalent labeling or cross-linking, incorporate labels to improve the ionization or dissociation of characteristic peptides in complex mixtures, and affinity-enrich various poorly abundant protein translational modifications (PTMs). A chemical modification reaction needs to be simple and efficient for use in proteomics analysis, and should be performed without any complicated process for preparing the labeling reagent.

View Article and Find Full Text PDF

Multiple Exciton Generation on Doped Wide-Band Semiconductor Photoanode with Hierarchical Quantum Structure.

Small

January 2025

Key Laboratory of Wide Bandgap Semiconductor Materials and Devices, School of Microelectronics, Xidian University, Xi'an, 710071, China.

The multiple exciton generation (MEG) effect, which produces multiple photo-generated charge carriers from a single high-energy photon absorption by a semiconductor with a narrow bandgap, has the potential to revolutionize photovoltaic, photoelectric detection, and other technologies. Here, this work finds that the surface carbon-modified wide-bandgap photoanode with hierarchical quantum structure can drive a photoelectrochemical reaction with a quantum efficiency exceeding 145% by the first time. More studies reveal that the presence of the MEG effect in the MEG-CdS photoanode is attributed to the formation of high-quality surface C-modified CdS quantum nanosheets on CdS bulk film by in situ, this hierarchical quantum structure leads to quantum confinement effects that increase effective Coulomb interaction for driving MEG and decrease competition for thermal exciton cooling.

View Article and Find Full Text PDF

Objective: Timing calibration is essential for positron emission tomography (PET) system as it enhances timing resolution to improve image quality. Traditionally, positron sources are employed for timing calibration. However, the photons emitted by these sources travel in opposite directions, necessitating that positrons annihilate at multiple locations to collect coincidence data across a greater number of lines of response (LORs).

View Article and Find Full Text PDF

Understanding how structural modifications affect the photophysics of organic linkers is crucial for their integration into metal-organic frameworks (MOFs) for light-driven applications. This study explores the impact of varying the amine functional group position on two terephthalic acid derivatives─linker and linker ─by investigating their photophysics through a combination of steady-state and ultrafast laser spectroscopy and time-dependent density functional theory (TD-DFT) calculations. With tetrahydrofuran as the solvent, time-correlated single-photon counting revealed a 2-fold increase in the S excited-state lifetime of the molecule with the amine group at the meta position compared with that of the molecule with the amine group at the ortho position.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!