A sensitive analytical method was developed for individual analyses of D- and L-glyceric acids by chiral derivatization - liquid chromatography-tandem mass spectrometry. To elucidate rapid and efficient optimization for derivatization we newly introduced a concept of design of experiments (DOE). The optimization of major 5 factors in the derivatization could be predicted with only 28 measurements. By applying DOE to optimization, the yields of desired derivatives increased five-fold against before optimization.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.2116/analsci.19N013 | DOI Listing |