Objective: To evaluate the benefits of adaptive imaging with automatic correction compared to periodic surveillance strategies with either manual or automatic correction.

Methods: Using Calypso trajectories from 54 patients with prostate cancer at 2 institutions, we simulated 5-field intensity-modulated radiation therapy and dual-arc volumetric-modulated arc therapy with periodic imaging at various frequencies and with continuous adaptive imaging, respectively. With manual/automatic correction, we assumed there was a 30/1 second delay after imaging to determine and apply couch shift. For adaptive imaging, real-time "dose-free" cine-MV images during beam delivery are used in conjunction with online-updated motion pattern information to estimate 3D displacement. Simultaneous MV-kV imaging is only used to confirm the estimated overthreshold motion and calculate couch shift, hence very low additional patient dose from kV imaging.

Results: Without intrafraction intervention, the prostates could on average have moved out of a 3-mm margin for ∼20% of the beam-on time after setup imaging in current clinical situation. If the time interval from the setup imaging to beam-on can be reduced to only 30 seconds, the mean over-3 mm percentage can be reduced to ∼7%. For intensity-modulated radiation therapy simulation, with manual correction, 110 and 70 seconds imaging periods both reduced the mean over-3 mm time to ∼4%. Automatic correction could give another 1% to 2% improvement. However, with either manual or automatic correction, the maximum patient-specific over-3 mm time was still relatively high (from 6.4% to 12.6%) and those patients are actually clinically most important. In contrast, adaptive imaging with automatic intervention significantly reduced the mean percentage to 0.6% and the maximum to 2.7% and averagely only ∼1 kV image and ∼1 couch shift were needed per fraction. The results of volumetric-modulated arc therapy simulation show a similar trend to that of intensity-modulated radiation therapy.

Conclusions: Adaptive continuous monitoring with automatic motion compensation is more beneficial than periodic imaging surveillance at similar or even less imaging dose.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6558533PMC
http://dx.doi.org/10.1177/1533033819844489DOI Listing

Publication Analysis

Top Keywords

adaptive imaging
20
imaging
12
automatic correction
12
intensity-modulated radiation
12
couch shift
12
periodic surveillance
8
prostate cancer
8
imaging automatic
8
manual automatic
8
radiation therapy
8

Similar Publications

Measurement of biomechanical properties of transversely isotropic biological tissue using traveling wave expansion.

Med Image Anal

January 2025

School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai 200030, China; Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai 200040, China; National Engineering Research Center of Advanced Magnetic Resonance Technologies for Diagnosis and Therapy (NERC-AMRT), Shanghai Jiao Tong University, Shanghai 200040, China; Department of Radiology, Ruijin Hospital affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai 200025, China. Electronic address:

The anisotropic mechanical properties of fiber-embedded biological tissues are essential for understanding their development, aging, disease progression, and response to therapy. However, accurate and fast assessment of mechanical anisotropy in vivo using elastography remains challenging. To address the dilemma of achieving both accuracy and efficiency in this inverse problem involving complex wave equations, we propose a computational framework that utilizes the traveling wave expansion model.

View Article and Find Full Text PDF

Investigating muscle architecture in static and dynamic conditions is essential to understand muscle function and muscle adaptations. Muscle architecture analysis, primarily through extended field-of-view ultrasound imaging, offers high reliability at rest but faces limitations during dynamic conditions. Traditional methods often involve "best fitting" straight lines to track muscle fascicles, leading to possible errors, especially with longer fascicles or those with nonlinear paths.

View Article and Find Full Text PDF

Therapeutic hurdles persist in the fight against lung cancer, although it is a leading cause of cancer-related deaths worldwide. Results are still not up to par, even with the best efforts of conventional medicine, thus new avenues of investigation are required. Examining how immunotherapy, precision medicine, and AI are being used to manage lung cancer, this review shows how these tools can change the game for patients and increase their chances of survival.

View Article and Find Full Text PDF

Background: Spinal cord injury/disease (SCI/D) profoundly affects both sexuality and urinary function. Catheterization is often necessary to manage bladder voiding and it can interfere with sexual activity.

Aim: We aim to investigate the effect of the bladder evacuation method on sexual activity in women with chronic SCI/D.

View Article and Find Full Text PDF

Echocardiography of the right heart in pulmonary arterial hypertension: insights from the ULTRA RIGHT VALUE study.

Eur Heart J Imaging Methods Pract

January 2025

Department of Clinical Internal, Anesthesiological and Cardiovascular Sciences, Sapienza University of Rome, Viale del Policlinico 155, Rome 00161, Italy.

Aims: Outcome in pulmonary arterial hypertension (PAH) is determined by right ventricular (RV) function adaptation to increased afterload. Echocardiography is easily available to assist bedside evaluation of the RV. However, no agreement exists about the feasibility and most relevant measurements.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!