Objective: The devices to be used in clinical practice should be tested for accuracy and should be validated through the validation protocol. The aim of this study was to determine the accuracy of the Meditech-ABPM-06 ambulatory blood pressure monitor.

Materials And Methods: The test was carried out in accordance with the protocol established by the European Society of Hypertension (ESH-IP), the British Hypertension Society, and the ANSI/AAMI/ISO 81060-2-2013 protocols.

Results: In the European Society of Hypertension validation procedure the subjects with two or three of the absolute differences between observer and device SBP/DBP measurements within 5 mmHg were 31/31. The average device-observer difference was -0.4 ± 3.9 mmHg for SBP and -0.4 ± 2.8 mmHg for DBP. In the British Hypertension Society validation procedure the mean differences of the test device and observer readings were -1.0 ± 4.6 (systolic) and -1.1 ± 4.5 (diastolic). In the ANSI/AAMI/ISO 81060-2-2013 validation procedure (criterion 1) the mean ± SD of the differences between the test device and reference BP was -0.5 ± 4.0/-0.3 ± 4.5 mmHg (systolic/diastolic). The two criteria of the ANSI/AAMI/ISO were fulfilled.

Conclusion: The Meditech ABPM-06 blood pressure measure device met the requirements specified by the British Hypertension Society, the European Society of Hypertension International Protocol and the AAMI, and therefore was considered acceptable to clinical use with a Grade of A/A.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MBP.0000000000000385DOI Listing

Publication Analysis

Top Keywords

european society
16
society hypertension
16
british hypertension
16
hypertension society
16
blood pressure
12
validation procedure
12
meditech abpm-06
8
ambulatory blood
8
society
8
hypertension
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!