Purpose: To evaluate the performance of an independent recalculation and compare it against current measurement-based patient specific intensity-modulated radiation therapy (IMRT) quality assurance (QA) in predicting unacceptable phantom results as measured by the Imaging and Radiation Oncology Core (IROC).

Methods: When institutions irradiate the IROC head and neck IMRT phantom, they are also asked to submit their internal IMRT QA results. Separately from this, IROC has previously created reference beam models on the Mobius3D platform to independently recalculate phantom results based on the institution's DICOM plan data. The ability of the institutions' IMRT QA to predict the IROC phantom result was compared against the independent recalculation for 339 phantom results collected since 2012. This was done to determine the ability of these systems to detect failing phantom results (i.e., large errors) as well as poor phantom results (i.e., modest errors). Sensitivity and specificity were evaluated using common clinical thresholds, and receiver operator characteristic (ROC) curves were used to compare across different thresholds.

Results: Overall, based on common clinical criteria, the independent recalculation was 12 times more sensitive at detecting unacceptable (failing) IROC phantom results than clinical measurement-based IMRT QA. The recalculation was superior, in head-to-head comparison, to the EPID, ArcCheck, and MapCheck devices. The superiority of the recalculation vs these array-based measurements persisted under ROC analysis as the recalculation curve had a greater area under it and was always above that for these measurement devices. For detecting modest errors (poor phantom results rather than failing phantom results), neither the recalculation nor measurement-based IMRT QA performed well.

Conclusions: A simple recalculation outperformed current measurement-based IMRT QA methods at detecting unacceptable plans. These findings highlight the value of an independent recalculation, and raise further questions about the current standard of measurement-based IMRT QA.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6692191PMC
http://dx.doi.org/10.1002/mp.13638DOI Listing

Publication Analysis

Top Keywords

independent recalculation
20
measurement-based imrt
20
detecting unacceptable
12
phantom
10
imrt
9
recalculation
9
imrt methods
8
methods detecting
8
unacceptable plans
8
current measurement-based
8

Similar Publications

Background: Modern radiation therapy techniques, such as intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT), use complex fluence modulation strategies to achieve optimal patient dose distribution. Ensuring their accuracy necessitates rigorous patient-specific quality assurance (PSQA), traditionally done through pretreatment measurements with detector arrays. While effective, these methods are labor-intensive and time-consuming.

View Article and Find Full Text PDF

Unaccounted clustering assumptions still compromise inferences in cluster randomized trials in orthodontic research.

Korean J Orthod

November 2024

Clinic of Orthodontics and Pediatric Dentistry, Center of Dental Medicine, University of Zurich, Zurich, Switzerland.

Objective: This meta-epidemiological study aimed to determine whether optimal sample size calculation was applied in orthodontic cluster randomized trials (CRTs).

Methods: Orthodontic randomized clinical trials with a cluster design, published between January 1, 2017 to December 31, 2023, in leading orthodontic journals were sourced. Study selection was undertaken by two independent authors.

View Article and Find Full Text PDF

Aims: Reporting of hospitalization cost data for the National Health Service in England evolved substantially in the past decades which may have led to inconsistency in capturing the average cost of hospitalizations in appraisals of new health technologies. Our aim was to systematically review the estimation of hospitalization costs in appraisals for the National Institute for Health and Care Excellence (NICE) and the Joint Committee on Vaccination and Immunisation (JCVI), to identify potential heterogeneity and its drivers. A case study was used to explore the impact of alternative cost estimates on economic evaluations.

View Article and Find Full Text PDF

Cytokine levels in the severity of falciparum malaria: An umbrella review.

Acta Trop

December 2024

College of Medicine and Dentistry, James Cook University, Queensland, Australia. Electronic address:

This study aimed to synthesise evidence comparing the levels of cytokines in severe falciparum malaria with those in uncomplicated malaria from available systematic reviews and meta- analyses. Relevant individual meta-analyses were searched in PubMed, Ovid, and Google Scholar, following the selection criteria specified for this umbrella review. The AMSTAR-2 tool was applied to grade the quality of the meta-analyses identified.

View Article and Find Full Text PDF

Background And Purpose: In vivo dosimetry is not standard in brachytherapy and some errors go undetected. The aim of this study was to evaluate the accuracy of multi-channel vaginal cylinder pulsed dose-rate brachytherapy using in vivo dosimetry.

Materials And Methods: In vivo dosimetry data was collected during the years 2019-2022 for 22 patients (32 fractions) receiving multi-channel cylinder pulsed dose-rate brachytherapy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!