Relative efficiency of Gafchromic EBT3 and MD-V3 films exposed to low-energy photons and its influence on the energy dependence.

Phys Med

Instituto de Física, Universidad Nacional Autónoma de México, 04510 Coyoacan, Mexico City, Mexico.

Published: May 2019

AI Article Synopsis

  • The energy dependence of Gafchromic films when exposed to low-energy photons is influenced by the absorbed dose, as highlighted in studies using a relative-response (R) measurement that doesn't account for the film’s sensitive area.
  • Research on Gafchromic EBT3 and MD-V3 films demonstrated that a new measure called relative-efficiency (RE) shows Co gamma rays require significantly higher absorbed doses compared to 20 kV x-rays to achieve the same optical density (netOD).
  • Findings suggest that a new factor (f), which considers energy-dependence influenced by film model, depth, and phantom material, could better evaluate film performance than the traditional relative-response methodology.

Article Abstract

Energy-dependence of Gafchromic films exposed to low-energy photons has been reported to be a function of absorbed-dose. However, these studies are based on a relative-response, R, which considers the absorbed-dose in water and not within the film sensitive-volume. This work investigated the relative-efficiency, RE, (ratio of absorbed-dose required to produce the same net optical density (netOD) by Co gamma and by x-ray) of Gafchromic EBT3 and MD-V3 films exposed to five x-ray beams from 20 kV to 160 kV and Co gamma rays. A factor that accounts for the energy-dependence, f, based on RE, phantom-material and depth at which the films are placed during irradiation was used to remove the influence of absorbed dose. Values of RE indicated that the absorbed dose from Co gamma rays needs to be 4 and 3 times larger than those from 20 kV x-rays to produce the same netOD within the EBT3 and MD-V3 sensitive volumes, respectively. Thus, saturation could help explain why Gafchromic films show under-response to very low doses from low-energy photon beams, regardless of film model. Furthermore, RE, was found to be nearly independent of netOD and colour-channels. Consequently, f is independent of the absorbed dose and colour-channels. In contrast, besides the variation with the photon energy, f varied with film model, depth and phantom material used during the irradiation. Thus, the results suggest that f is a more reliable wide-ranging parameter for evaluating the degree of energy-dependence of the film rather than the relative-response method commonly considered.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2019.04.007DOI Listing

Publication Analysis

Top Keywords

ebt3 md-v3
12
films exposed
12
absorbed dose
12
gafchromic ebt3
8
md-v3 films
8
exposed low-energy
8
low-energy photons
8
gafchromic films
8
gamma rays
8
film model
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!