Background: Although ultrashort echo time (UTE) sequences allow excellent assessment of lung parenchyma, image quality remains lower than that of computed tomography (CT).

Purpose: To investigate a high-frequency noninvasive ventilation (HF-NIV) technique allowing a stabilized inspiration and to compare image quality with current dedicated MR sequences.

Study Type: Prospective.

Population: Ten healthy volunteers.

Field Strength/sequence: 3D radial UTE sequence at 1.5T.

Assessment: UTE-HF-NIV sequence was compared with UTE-free-breathing (UTE-FB), reconstructed at end expiration (UTE-Exp) and average (UTE-Avg), and breath-hold VIBE sequences. The distance from lung apex to the dome of the right hemidiaphragm was measured. Visual assessment of the visibility and sharpness of normal anatomical structures was carried out. Dedicated software also quantitatively evaluated vessel-lung and right lung-liver interface sharpness. Apparent signal ratio (Sr) and contrast ratios (Cr) were quantitatively evaluated.

Statistical Tests: Wilcoxon signed rank test for visual scores, paired t-test for continuous variables, significance at P < 0.05.

Results: The distance between apex and the right hemidiaphragmatic dome was significantly larger (P < 0.001) with UTE-HF-NIV compared with UTE-FB and VIBE acquisitions. Vessel and airway visibility had identical median visual scores with all UTE methods. Median visual scores for sharpness of vessels and airways were significantly higher (P < 0.001) with HF-NIV (vessels = 3; airways = 2) than in UTE-FB (vessels = 2; airways = 1) and VIBE (vessels = 1; airways = 1). Software-based vessel sharpness evaluation resulted in larger values in 8/10 volunteers with UTE-HF-NIV (67.3 ± 9.8) compared with UTE-Avg (62.3 ± 12.6) but the average difference was not significant (P = 0.28). The sharpness of the lung-liver interface was significantly higher (P < 0.001) with HF-NIV (17.3 ± 5.3) compared with UTE-Avg (14.1 ± 3.9). Significantly higher values (P < 0.01) of Sr and Cr were observed with UTE-HF-NIV compared with UTE-FB and VIBE.

Data Conclusion: HF-NIV allowing acquisition at full inspiration significantly improves image quality for lung imaging. This could offer the option to alternate some follow-up CT studies by using this technique.

Level Of Evidence: 2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;50:1789-1797.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6900075PMC
http://dx.doi.org/10.1002/jmri.26808DOI Listing

Publication Analysis

Top Keywords

ultrashort echo
8
echo time
8
high-frequency noninvasive
8
noninvasive ventilation
8
image quality
8
time imaging
4
imaging lungs
4
lungs high-frequency
4
ventilation approach
4
approach lung
4

Similar Publications

It is known that ultrashort echo time (UTE) magnetic resonance imaging (MRI) sequences can detect signals from water protons but not collagen protons in short T2 species such as cortical bone and tendons. However, whether collagen protons are visible with the zero echo time (ZTE) MRI sequence is still unclear. In this study, we investigated the potential of the ZTE MRI sequence on a clinical 3T scanner to directly image collagen protons via DO exchange and freeze-drying experiments.

View Article and Find Full Text PDF

Background: Early detection of changes in lumbar intervertebral disc degeneration (IVDD) has great clinical significance. T1rho and UTE-T2* relaxometry are capable of providing information about biochemical changes in IVDD. However, they have not been previously analyzed and compared directly in the same patients.

View Article and Find Full Text PDF

Utilizing a multi-task deep learning framework, this study generated synthetic CT (sCT) images from a limited dataset of Ultrashort echo time (UTE) MRI for transcranial focused ultrasound (tFUS) planning. A 3D Transformer U-Net was employed to produce sCT images that closely replicated actual CT scans, demonstrated by an average Dice coefficient of 0.868 for morphological accuracy.

View Article and Find Full Text PDF

Introduction: Restricted fetal and neonatal growth is a known risk factor for bronchopulmonary dysplasia (BPD) in premature infants. However, the impact of nutrition and infant growth specifically on lung growth in BPD in unknown. Moreover, whether all lung growth in BPD is beneficial is unclear.

View Article and Find Full Text PDF

Aims: The aim of this study was to develop an ultra-short echo time 3D magnetic resonance imaging (MRI) method for imaging subacute myocardial infarction (MI) quantitatively and in an accelerated way. Here, we present novel 3D T- and T -weighted Multi-Band SWeep Imaging with Fourier Transform and Compressed Sensing (MB-SWIFT-CS) imaging of subacute MI in mice hearts .

Methods And Results: Relaxation time-weighted and under-sampled 3D MB-SWIFT-CS MRI were tested with manganese chloride (MnCl) phantom and mice MI model.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!