Aims: The proximal optimization technique (POT) in coronary bifurcation stenting improves apposition and side-branch obstruction. The POT balloon should be positioned with the distal radio-opaque marker at the carina cut plane. However, the real impact of positioning remains unknown.

Methods And Results: Synergy™ stents (Boston Scientific, USA) were implanted on left-main fractal bench models. Initial POT was performed in 3 positions according to distal shoulder position (loss of balloon parallelism) relative to the carina cut plane (n = 5/group): i) "proximal", 1 mm before carina; ii) "medium", just at carina; iii) "distal", 1 mm after carina. Results were quantified on 2D- and 3D-OCT. Compared to implantation, initial POT improved malapposition in all positions ("proximal": 61.5 ± 1.4% vs. 5.1 ± 2.7%; "medium": 60.2 ± 2.4% vs. 1.3 ± 0.6%; "distal": 60.5 ± 2.9% vs. 1.1 ± 1.8%, p < 0.05). However, residual malapposition was greater in "proximal" position (p < 0.05). "Proximal", unlike "medium" or "distal" POT, also failed to improve side-branch obstruction. Conversely, "distal" POT significantly overstretched the main-branch ostium, with stent/artery ratio 1.22 ± 0.04 vs. 1.11 ± 0.07 for "medium" POT (p < 0.05).

Conclusion: Shoulder positioning is essential to optimize the mechanical benefit of POT without main-branch overstretch (too distal position). Experimentally, the best position is just at the carina cut plane ("medium").

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijcard.2019.05.041DOI Listing

Publication Analysis

Top Keywords

proximal optimization
8
carina cut
8
cut plane
8
initial pot
8
1 mm carina
8
carina
5
optimal balloon
4
balloon positioning
4
positioning proximal
4
optimization technique?
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!