Background: Most depth of anesthesia (DOA) monitors rely on the temporal characteristics of a single-channel electroencephalogram (EEG) and cannot provide spatial or connectivity information. Phase lag entropy (PLE) reflects DOA by calculating diverse connectivity from temporal patterns of phase relationships. The aim of this study was to compare the performance of PLE and bispectral index (BIS) monitors for assessing DOA during anesthesia induction, nerve integrity monitoring (NIM), and anesthesia emergence.

Methods: Thirty-five patients undergoing elective thyroid surgery with recurrent laryngeal nerve NIM received propofol and remifentanil via target-controlled infusion. After applying PLE and BIS monitors, propofol infusion was initiated at a calculated effect site concentration (Ce) of 2 μg/mL and then increased in 1-μg/mL Ce increments. After propofol Ce reached 5 μ/mL, a remifentanil infusion was begun, and anesthesia induction was considered complete. During NIM, PLE and BIS values were compared at a specific time points from platysma muscle exposure to subcutaneous tissue closure. PLE and BIS values were recorded continuously from preanesthetic state to full recovery of orientation; bias and limits of agreement between monitors were calculated.

Results: PLE and BIS values decreased progressively with increasing propofol Ce during anesthetic induction and increased by stages during emergence. The prediction probabilities of PLE and BIS for detecting propofol Ce changes were 0.750 and 0.756, respectively, during induction and 0.749 and 0.746, respectively, during emergence. No aberrant PLE or BIS values occurred during NIM. Correlation coefficients for BIS and PLE were 0.98 and 0.92 during induction and emergence, respectively. PLE values were significantly higher than BIS values at full recovery of orientation. Estimated bias between monitors was -4.16 ± 8.7, and 95% limits of agreement were -21.21 to 12.89.

Conclusion: PLE is a reasonable alternative to BIS for evaluating consciousness and DOA during general anesthesia and during NIM.

Trial Registration: Clinical Research Information Service Identifier: KCT0003490.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6535403PMC
http://dx.doi.org/10.3346/jkms.2019.34.e151DOI Listing

Publication Analysis

Top Keywords

ple bis
24
bis values
20
ple
11
bis
10
phase lag
8
lag entropy
8
patients undergoing
8
thyroid surgery
8
nerve integrity
8
integrity monitoring
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!