Introduction: Headless compressions screws are the most implanted devices for scaphoid fractures and nonunions. For cases when screw osteosynthesis is not possible, a special locking plate for scaphoid reconstruction has been developed. The purpose of this study was to evaluate the safety and practicability of this device for difficult scaphoid pathologies.
Materials And Methods: Between March 2010 and December 2014, 20 patients (age range 16-59 years) were treated with scaphoid locking plate osteosynthesis. In 17 cases it was due to scaphoid nonunion or delayed union and in three cases to treat a complex multi-fragmentary fracture of the scaphoid. Most of the initial fractures were located either in the proximal third (n = 9) or the middle third (n = 8) of the scaphoid.
Results: Mean follow-up was 14.6 ± 8.9 months (range 2-30 months). All three scaphoid fractures (100%) showed bony healing in the CT scan after 2.7 ± 0.6 months. 15 of 17 (88.2%) patients with scaphoid nonunion demonstrated bony healing in the latest CT scan at an average of 6.2 ± 8.1 months (range 2-11 months) after scaphoid reconstruction. Range of motion (extension/flexion) was 104° ± 18.4° (range 80°-150°) and about one third less than the unaffected side. The average grip strength averaged 38.2 kg on the operated side and 44.1 kg on the unaffected side after surgery. 13 plates (65%) had to be removed due to impaction of the plate or protrusion of the screws.
Conclusions: This new locking device for scaphoid reconstruction seems to be a safe, useful and reliable tool in the treatment of difficult nonunions or multi-fragmentary scaphoid fractures. The practicability is convincing and satisfying fusion rates can be accomplished. However, most patients require hardware removal. We recommend using this plate as a rescue option when a stable osteosynthesis is necessary for the healing process and screw fixation has already failed or is not possible.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00402-019-03196-6 | DOI Listing |
Cureus
December 2024
Oral and Maxillofacial Surgery, Nagoya Tokushukai General Hospital, Kasugai, JPN.
Sternal fractures resulting from blunt chest trauma often present unique surgical challenges. While conservative management is common, cases with significant displacement, delayed union, or painful dyspnea may require surgical intervention to improve structural stability and relieve symptoms. Here, we report the case of a 46-year-old man who sustained a displaced sternal fracture in a motor vehicle accident.
View Article and Find Full Text PDFJ Hand Microsurg
January 2025
Orthopaedics and Traumatology Unit, Cattinara Hospital - ASUGI, Department of Medical, Surgical and Life Sciences, Trieste University, Trieste, Italy.
Background: Aim of the present paper is to report the preliminary results of CAD-CAM (Computer-Aided Design - Computer-Aided Manufacturing) technology application to distal femur nonunion treatment with free fibula flap, custom made medial plating and maintenance of a stable lateral locking plate.
Methods: Two cases of distal femur nonunion that occurred after lateral locking plating were treated and prospectively followed-up. Surgical planning followed the same preoperative protocol adopted for mandibular CAD-CAM reconstruction.
Sci Rep
January 2025
Department of Orthopaedic Trauma, The Second Affiliated Hospital of Dalian Medical University, No. 467 Zhongshan Road, Shahekou District, Dalian, 116027, Liaoning, China.
Anterior cervical interbody fusion (ACDF) has become a classic surgical procedure for the treatment of cervical degenerative diseases, and various interbody cages are widely used in this procedure. We used 3D printing technology to produce a new type of plate-locking cage, anticipating to achieve high fusion rate with the high biomechanical stability. This study is to compare the biomechanical characteristics between a newly designed interbody cage and a conventional Zero-profile cage during ACDF using finite element analysis.
View Article and Find Full Text PDFActa Orthop Traumatol Turc
December 2024
Department of Orthopaedics and Traumatology, Izmir Katip Celebi University, Ataturk Training and Research Hospital, Izmir, Türkiye.
Objective: The aim of this study was to evaluate whether the locking femoral neck plate (LFNP) can be an alternative fixation method to the cannulated screws with a medial buttress plate. For this purpose, we compared biomechanically the LFNP and cannulated screws with or without a medial buttress plate in Pauwels type 3 femoral neck fractures.
Methods: A vertical fracture model was created at an 80-degree angle to the femoral neck in 28 synthetic bone models.
JSES Rev Rep Tech
February 2025
Department of Orthopaedics, Krishna Institute of Medical Sciences, Secunderabad, Telanagana, India.
Hypothesis: Proximal humerus fractures present a treatment challenge due to varied fracture configurations and a lack of consensus on optimal management. Locking plate designs offer promising solutions, yet technical guidelines for successful outcomes remain elusive. Complications are common, with fixation-related failures often attributed to varus collapse.