Objectives: The VIVA (Valve in Valve) trial was designed to systematically and prospectively collect data regarding the use of transcatheter aortic valve replacement in patients with failing surgical aortic bioprostheses at high-risk for reoperation.
Background: Surgical aortic valve replacement has been the standard of care in symptomatic patients with aortic valve disease. However, bioprosthetic valves degenerate over time, requiring redo surgery.
Methods: VIVA is an international, observational, single-arm, postmarket study conducted at 23 sites that enrolled 202 patients with symptomatic degeneration of an aortic bioprosthesis eligible for elective treatment with a CoreValve or Evolut R self-expanding transcatheter aortic valve.
Results: Patients were elderly (mean age 79.9 years), 47.5% were men, and they had a mean Society of Thoracic Surgeons score of 6.6%. Although 41.8% of patients had surgical bioprostheses with labeled size ≤21 mm, valve hemodynamic parameters were markedly improved from baseline (mean aortic valve gradient 35.0 ± 16.3 mm Hg) to discharge (17.5 ± 8.6 mm Hg) and were sustained at 1 year (15.5 ± 7.5 mm Hg). At 1 year, total aortic regurgitation greater than mild was measured in 1.1% of patients. Clinical outcomes at 30 days demonstrated low mortality (2.5%), no disabling strokes, a 0.5% rate of acute kidney injury, and an 8.0% rate of new pacemaker implantation. At 1 year, the mortality rate remained low (8.8%), with 1 disabling stroke (0.6%). Five patients (2.5%) experienced coronary artery obstructions, 3 during and 1 immediately after the procedure and 1 several months later.
Conclusions: Degenerated surgical bioprostheses can be safely treated with the CoreValve or Evolut R platform using the catheter-based valve-in-valve procedure. Excellent 1-year clinical and hemodynamic outcomes were achieved in this real-world patient population. (CoreValve VIVA Study Evaluation of the Clinical Outcomes of CoreValve in Degenerative Surgical Aortic Bioprosthesis; NCT02209298).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcin.2019.02.029 | DOI Listing |
The zebrafish is a valuable model organism for studying cardiac development and diseases due to its many shared aspects of genetics and anatomy with humans and ease of experimental manipulations. Computational fluid-structure interaction (FSI) simulations are an efficient and highly controllable means to study the function of cardiac valves in development and diseases. Due to their small scales, little is known about the mechanical properties of zebrafish cardiac valves, limiting existing computational studies of zebrafish valves and their interaction with blood.
View Article and Find Full Text PDFAnatol J Cardiol
January 2025
Department of Cardiothoracic Surgery Research, Lankenau Institute for Medical Research, Wynnewood, Pennsylvania, USA ; Department of Cardiothoracic Surgery, Lankenau Heart Institute, Main Line Health Wynnewood, Pennsylvania, USA.
Background: To evaluate the clinical outcomes of valve-in-valve transcatheter aortic valve replacement (ViV TAVR) with newer-generation self-expanding Evolut valves according to the size of the failed surgical bioprosthesis.
Methods: This single-center retrospective study evaluated consecutive patients undergoing ViV TAVR with the Evolut Pro/Pro+/Fx between 2018 and 2022. These patients were compared based on the true internal diameter (ID) of the failed bioprosthesis, specifically ≤19 mm (small group) vs.
Ann Noninvasive Electrocardiol
January 2025
Division of Cardiology, Department of Medicine, Brigham and Women's Hospital, Boston, Massachusetts, USA.
Background: Electrocardiograms (EKGs) are routinely performed in pregnant patients with pre-existing cardiovascular disease. However, in pregnant patients with congenital heart disease (CHD), EKG changes during gestation have not been explored.
Methods: We performed a retrospective study of pregnant patients with CHD enrolled in the STORCC initiative.
Gen Thorac Cardiovasc Surg Cases
January 2025
Department of Thoracic and Cardiovascular Surgery, Wakayama Medical University, 811-1 Kimiidera, Wakayama City, Wakayama, 641-8509, Japan.
Patients with coronary artery disease undergoing trans-catheter aortic valve implantation (TAVI) often receive TAVI alone. However, in cases of severe coronary lesions or anticipated difficulty in coronary access post-TAVI, percutaneous coronary intervention or coronary artery bypass grafting may be necessary. We performed simultaneous gastroepiploic artery to posterior descending artery bypass and TAVI in two patients with severe calcification of the right coronary artery ostium which is unsuitable for percutaneous intervention.
View Article and Find Full Text PDFCardiovasc Interv Ther
January 2025
Division of Cardiology, Department of Medicine, Kindai University Faculty of Medicine, 377-2 Ohno-Higashi, Osakasayama, Osaka, 589-8511, Japan.
Transcatheter aortic valve implantation (TAVI) using the NAVITOR system has been relatively underreported due to its recent introduction in Japan. This study aimed to assess the short-term outcomes of TAVI with the NAVITOR in real-world clinical practice. Patients with severe aortic stenosis who underwent TAVI using the NAVITOR system at our institution between December 2022 and December 2023 were prospectively enrolled.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!