  • Photoplethysmographic imaging (PPG) is used to measure heart rate (HR), but skin pigmentation may affect its accuracy, especially during exercise.
  • A study evaluated the Apple Watch's PPG sensor against a Polar monitor across different exercise intensities in males with various skin types (II, III, IV) to determine if skin type impacts HR measurement accuracy.
  • Findings revealed strong agreement between devices (correlation 0.96-0.99), with minimal influence from skin type, suggesting the Apple Watch can reliably monitor HR during cycling regardless of skin color, which is relevant for exercise intensity management.

Article Abstract

Photoplethysmographic imaging (PPG) is currently used to measure heart rate (HR) and the accuracy of PPG can be influenced by pigmentation of the skin; however, the effects of skin color-related artifacts on PPG during exercise remain unclear. This study aimed to assess the agreement between the Apple Watch photoplethysmography sensor and a criterion, for measuring heart rate across a range of intensities during exercise and to determine the influence of skin type on the accuracy of the measure. Forty-five males (20-43 y) completed the Fitzpatrick Skin Scale and were classified into three different skin type groups: a) types II (n = 15), III (n = 15) and IV (n = 15). Participants performed a graded incremental cycle-ergometer test while simultaneously wearing the Apple Watch and a Polar monitor as a criterion measure. Data from both devices were collected in 5-s epochs. Correlations between devices were very good (0.96-0.99 [95%CI: 0.94 to 0.99]). Significant differences were observed between skin types II and III when the intensity of the exercise was increased, albeit with trivial to small effect sizes (ES: 0.05 to 0.28). All significant differences corresponded to <2% of relative difference between both devices. Bland-Altman analyses showed a trivial but systematic underestimation of HR in the Apple Watch compared to Polar for all skin types during exercise. In conclusion, the Apple Watch accurately measures HR when cycling at different intensities and certain types of skin seem not to influence these measures, which may have important implications for controlling the intensity of exercise.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10916-019-1325-2DOI Listing

Publication Analysis

Top Keywords

skin type
12
heart rate
12
apple watch
12
influence skin
8
skin
7
pilot study
4
study assessing
4
assessing influence
4
type heart
4
rate measurements
4

Similar Publications

Introduction: The reverse sural flap (RSF) is a random-type, pedicled flap based on sural artery perforators indicated for traumatic lower-extremity wounds. The RSF has demonstrated comparable results to free flap placement in the adult population for reconstruction of distal third defects, but few reports describe its application and outcomes in the pediatric population.

Methods: We investigated RSF application in pediatric patients (<18 years of age) through systematic review and meta-analysis.

View Article and Find Full Text PDF

Appropriate surgical management in skin and soft tissue infections.

Curr Opin Infect Dis

January 2025

Department of Surgery, Klinik Donaustadt, Vienna Healthcare Group, Austria.

Purpose Of Review: To present standards and recent technical innovations in the surgical management of skin and soft tissue infections (SSTI).

Recent Findings: SSTI are a frequent cause of presentation in the acute care setting. They can range from simple and uncomplicated to severe and necrotizing infections.

View Article and Find Full Text PDF

Background: Egg allergy is among the most common food allergies in children, significantly affecting the dietary habits and quality of life of both the affected children and their families. This study aims to assess the clinical role of the Basophil Activation Test (BAT) in children with egg allergy and to evaluate its diagnostic accuracy in comparison to other tests.

Methods: The study included 46 children with egg allergy.

View Article and Find Full Text PDF

sp. nov., isolated from human epidermis.

Int J Syst Evol Microbiol

January 2025

Department of Bio Health Science, Changwon National University, Changwon, Gyeongnam 51140, Republic of Korea.

Five pink-pigmented bacterial strains, isolated from human skin and classified within the genus , were examined. Among them, four were identified as , while strain OT10 was deemed to be a potential novel species. Strain OT10 exhibited characteristics, such as Gram-stain-negative, oxidase positive, motile, strictly aerobic and rod shaped.

View Article and Find Full Text PDF

Summary: Palmoplantar keratoderma (PPK), characterised by excessive epidermal thickening of the skin on the palms and/or plantar surfaces of the feet, can be hereditary or acquired. Here, we report a case of a 53-year-old woman with a history of sub-optimally controlled diabetes mellitus presenting with fevers and decreased Glasgow Coma Scale (GCS) to a tertiary hospital. She was diagnosed with diabetic ketoacidosis (DKA), with blood glucose at 40 mmol/L and ketones at 7 mmol/L, in the setting of a methicillin-sensitive Staphylococcus aureus necrotising soft tissue back infection.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!