Objective: To evaluate the optimum percent on time for the most efficient lens fragment removal using long-pulse torsional ultrasound (US).
Design: In vitro laboratory study.
Methods: Porcine lens nuclei were incubated in formalin for 2 hours and then cut into 2 mm cubes. Phacoemulsification was performed using the Centurion Vision System and Infiniti OZil handpiece with the balanced tip. Vacuum was set at 500 mm Hg, aspiration rate at 50 mL/min, and intraocular pressure (IOP) at 50 mm Hg. Pulse rate was 26 pulses/second. Studied parameters were percent power: 60%, 80%, and 100%, and percent on times: 50%, 60%, 70%, and 80%. Efficiency was the total time for a cube to be emulsified. Chatter was the number of times the lens fragment bounced off the tip.
Results: There was no significant difference in efficiency between 50%, 60%, and 70 % on-time settings (p = 0.17 and 0.08, respectively); however, there was significant increase in efficiency when the on time was increased from 70% to 80% (p = 0.03). Increasing power from 60% to 100% showed a statistically significant efficiency increase (p = 0.001). There was no significant change in chatter with increasing on time; however, there was a statistically significant increase in chatter with every power level increase.
Conclusion: Increasing on-time percent does not improve efficiency under torsional long-pulse US. There is no significant change in chatter with increasing on-time percent. Increasing power increases efficiency despite chatter increase. Long-pulse US does not appear to influence torsional action in a clinically meaningful way.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcjo.2018.07.005 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!