Purpose: We aimed to compare the size-specific dose estimates (SSDE), computed tomography (CT) dose indices and image quality parameters of the chest CTs obtained with fast kilovoltage-switching (FKS) dual-energy (DE) CT versus those with single-energy (SE) CT.
Methods: Patients who had chest SECT within the last 6 months were prospectively scanned with chest FKS-DECT. Quantitative comparison was made by calculating the mean SSDE, CTDIvol, contrast, noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) for both acquisitions. Two radiologists evaluated the chest SECT and DECT images qualitatively blinded to the technique used. The paired Student's t test was utilized for comparing the quantitative and qualitative data. Inter- and intraobserver agreement were also assessed.
Results: A total of 42 patients were included. The mean SSDE, CTDIvol, contrast, noise, CNR, and SNR for SECT versus DECT were 12.7±2.2 mGy vs. 9.3±1.2 mGy (P = 0.001), 10.9±2.4 mGy vs. 8±1.2 mGy (P < 0.001), 211.9±44.7 vs. 216.3±59 (P = 0.350), 12.9±2.4 vs. 13.9±3.7 (P = 0.086), 13.5±5.2 vs. 13.3±8.4 (P = 0.548) and 12±3.5 vs. 11.5±3.4 (P = 0.774), respectively. Interobserver reproducibility was high for contrast, noise, CNR, and SNR (ICC = 0.89, 0.85, 0.93, and 0.82, respectively; all P < 0.05). Intraobserver reproducibility was high for contrast, noise, CNR, and SNR (ICC = 0.80, 0.77, 0.85, and 0.88, respectively; all P < 0.05).
Conclusion: The mean SSDE of the chest CTs obtained with FKS-DECT were 26.8% lower than those with SECT with significant difference for the objective assessment and there was no significant difference for the subjective assessment of the image qualities, in this series.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6521909 | PMC |
http://dx.doi.org/10.5152/dir.2019.18412 | DOI Listing |
Eur J Nucl Med Mol Imaging
January 2025
Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.
Purpose: The positron range effect can impair PET image quality of Gallium-68 (Ga). A positron range correction (PRC) can be applied to reduce this effect. In this study, the effect of a tissue-independent PRC for Ga was investigated on patient data.
View Article and Find Full Text PDFJ Comput Assist Tomogr
January 2025
Department of Radiology, College of Medicine, University of Florida, Gainesville, FL.
Purpose: The purpose of this work was to evaluate the image quality of a commercial CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO).
Methods: CT number, noise, contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectrum (NPS) were assessed using the ACR CT Accreditation phantom scanned with various acquisitions at 80 kV, 100 kV, 120 kV, and 135 kV, each with multiple CTDIvol values of 20 mGy, 40 mGy, and 65 mGy. Artifacts were evaluated in an anthropomorphic head phantom, a cadaver head, and in patient studies.
J Comput Assist Tomogr
January 2025
Department of Radiology, Affiliated Hospital of Shaanxi University of Chinese Medicine.
Objective: To explore the application of low-energy image in dual-energy spectral CT (DEsCT) combined with deep learning image reconstruction (DLIR) to improve inferior vena cava imaging.
Materials And Methods: Thirty patients with inferior vena cava syndrome underwent contrast-enhanced upper abdominal CT with routine dose, and the 40, 50, 60, 70, and 80 keV images in the delayed phase were first reconstructed with the ASiR-V40% algorithm. Image quality was evaluated both quantitatively [CT value, SD, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) for inferior vena cava] and qualitatively to select an optimal energy level with the best image quality.
J Comput Assist Tomogr
January 2025
Department of Radiology, the Affiliated Traditional Chinese Medicine Hospital, Southwest Medical University.
Background: With the widespread use of lumbar pedicle screws for internal fixation, the morphology of the screws and the surrounding tissues should be evaluated. The metal artifact reduction (MAR) technique can reduce the artifacts caused by pedicle screws, improve the quality of computed tomography (CT) images after pedicle fixation, and provide more imaging information to the clinic.
Purpose: To explore whether the MAR+ method, a projection-based algorithm for correcting metal artifacts through multiple iterative operations, can reduce metal artifacts and have an impact on the structure of the surrounding metal.
We demonstrate a widely spaced, stabilized, and self-referenced opto-electronic oscillator driven electro-optic modulator based optical frequency comb. Using an ultra-stable Fabry-Perot etalon as a stable reference, we simultaneously stabilize a CW laser and generate a low noise and stable RF oscillation used to drive an electro-optic comb. In such a manner, the Fabry-Perot etalon pins both the carrier-envelope-offset frequency ( ) and the repetition rate of the comb in place ( ), eliminating the need for an external RF oscillator.