: The aim of the study was to analyze the accuracy of impedance cardiography (ICG) for hemodynamic assessment in wheelchair rugby players during rest and exercise. : The study included 21 players (mean age 33.0 ± 5.4, 86% male) with posttraumatic tetraplegia. ECG, echocardiography, and gas exchange analysis during rest and exercise were used to obtain heart rate (HR), stroke volume (SV), and cardiac output (CO) for comparison with PhysioFlow®. : There was a good correlation between reference methods and ICG for HR, SV, CO at rest and CO at peak exercise (r = 0.69-0.77, < .001) and a very good correlation for peak HR (r = 0.91, < .0001). ICG overestimated SV at rest, CO at rest, and peak CO, which resulted in low intraclass correlation coefficients (ICC = 0.250 and 0.570). : ICG can serve as a good estimate of basic hemodynamic parameters during rest and exercise in wheelchair rugby players but overestimates stroke volume and cardiac output.