Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing.

Clin Oral Investig

Department of Prosthetic Dentistry and Dental Materials Sciences, Saarland University, Homburg Campus, Bldg. 71.2, 66421, Homburg, Saar, Germany.

Published: July 2019

Objective: Aim was to assess the influence of a standard autoclaving protocol on the retention of zirconia abutments glued on titanium bases in two-piece implant abutments.

Materials And Methods: Twenty-four CAD/CAM-generated zirconia abutments were adhesively cemented on prefabricated titanium bases. Before mechanical and thermal aging, the specimens were divided into two groups. Group 1 was subjected to a standard steam autoclaving protocol and Group 2 remained untreated (control). The tensile strength in all specimens was evaluated by a standardized pull-off test limited to a maximum force of 1000 N.

Results: Eleven samples in both groups could be subjected to pull-off testing. Ten samples in Group 1 and three samples in Group 2 failed, while all others reached the maximum pull-off force. This difference was statistically significant. The mean retention values for the failed samples were 694.53 ± 369.10 N in Group 1 and 890.78 ± 25.90 N in Group 2. This difference was not statistically significant.

Conclusions: A standard autoclaving protocol does not reduce detachment force of two-piece zirconia abutments.

Clinical Relevance: Clinical sterilization processes as recommended by regulatory authorities seem to be harmless to the structural integrity of two-piece zirconia implant abutments, at least with regard to the retention of the components.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00784-019-02926-9DOI Listing

Publication Analysis

Top Keywords

autoclaving protocol
16
standard autoclaving
12
two-piece zirconia
12
zirconia abutments
12
structural integrity
8
integrity two-piece
8
detachment force
8
titanium bases
8
samples group
8
difference statistically
8

Similar Publications

Since late 2021, outbreaks of highly pathogenic avian influenza virus have caused a record number of mortalities in wild birds, domestic poultry, and mammals in North America. Wetlands are plausible environmental reservoirs of avian influenza virus; however, the transmission and persistence of the virus in the aquatic environment are poorly understood. To explore environmental contamination with the avian influenza virus, a large-volume concentration method for detecting infectious avian influenza virus in waterbodies was developed.

View Article and Find Full Text PDF

Reusing healing abutments is common practice among clinicians; however, ensuring complete surface decontamination is crucial to avoid further complications. This study aimed to evaluate the cleaning potential of Er: YAG laser at different frequencies, as an adjunctive step prior to autoclave sterilization. Forty contaminated healing abutments were divided into four groups.

View Article and Find Full Text PDF

Disinfection practices in dental settings are fundamental to clinical safety, playing a pivotal role in preventing cross-infections and protecting the health of patients and healthcare professionals. This article examines the key components of effective disinfection, based on evidence-based protocols developed by international organizations such as the WHO and the U.S.

View Article and Find Full Text PDF

Background: Enterococcus faecalis (E. faecalis) is the most common microorganism responsible for post-endodontic infections, which can penetrate deep into dentinal tubules. This study aimed to compare the survival of E.

View Article and Find Full Text PDF

Surgical Management and Proposed Protocol for Iatrogenic Separation of Bur During Maxillofacial Surgery.

J Maxillofac Oral Surg

December 2024

Present Address: Department of Dental and Oral Maxillofacial Surgery, LHMC and Associated SSK and KSC Hospital, New Delhi, India.

Medical device separation or breakage is not an uncommon incident in routine maxillofacial surgery. The maxillofacial surgical procedures utilize various devices like piezosurgery tips, tungsten carbide burs, chisels and elevators. Instrument breakage or separation may be attributed to substandard grade, overused blunt instruments, repeated autoclaving, improper force application and overheating of instruments.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!